PERFUSION REPORTS

REPORT
059
DATE
1/5/2020
TITLE
Failure to remove CO2 effectively
DETAILS OF INCIDENT

There was a failure to remove CO2 with normal sweep gas flow rates. The patient had a calculated flow of 5LPM. Sweep gas flow rates of between 7.5-9 LPM were required to achieve a pCO2 of 5kPa. These high sweeps were required from the initiation of CPB and for the full duration of the procedure. pO2's were normal throughout.

The case was a CABGx3 with a target temperature of 34degrees Celsius. Total CPB time was 95mins.

The baseline ACT was 86 seconds. 10,000units of heparin were added to prime as per local unit protocol. 30,000 units of heparin were given pre-CPB to achieve an ACT of 466 pre-CPB. The ACT machine and cartridges used (Haemochron Signature Elite) had all been QC'd as per manufacturers recommendations. The Maquet HLM20 was used alongside rotaflow centrifugal head, electronic blender and Terumo CDI.

Baseline pre-op pCO2 was 5.4kPa. Ventilated pre-CPB pCO2 was 5.9kPa. The first ACT on CPB was 526, followed by an ACT of 443, where an additional 5000 units of heparin were administered. This returned an ACT value of 516. There was no CO2 flushing for the case and the outlet of oxygenator was checked for any obstructions.

Drugs administered throughout CPB included heparin, phenylephrine, cardioplegia, magnesium and lignocaine.

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

Getinge Quadrox oxygenator

Lot No: 70131464

ACTION TAKEN

The vapouriser was temporarily turned off and the scavenging removed to see if it made a difference. There was no change to pCO2. The gas blender was switched out and high sweeps were still required with the alternative Sechrist blender. The pO2's were within normal ranges for the duration of CPB. The oxygenator was carefully monitored throughout the case in case of the need for a changeout.

The surgeon was informed post CPB. The manufacturer was informed and the device was retained and returned unwashed to the manufacturer for investigation.

RESPONSE FROM MANUFACTURER

See attached

ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
089
1/9/21
Sechrist Gas Blender Sweep on ECMO
REPORT NO
088
1/9/21
Oxygenator High Pressure Events
REPORT NO
087
1/10/21
Leaking one way valve
REPORT NO
086
1/7/21
Electronic Gas Blender Fault
REPORT NO
085
1/2/21
Protamine Administration Issue
REPORT NO
084
1/6/21
Damaged roller pump lid
REPORT NO
083
1/6/21
IABP malfunction
REPORT NO
082
1/1/21
Accidental cardioplegia administration
REPORT NO
081
1/4/21
S5 Double Roller Pump incident
REPORT NO
080
1/5/21
Human factors issue
REPORT NO
079
1/6/21
Centrimag battery module failure
REPORT NO
078
1/6/21
Sampling manifold valve fault
REPORT NO
077
1/5/21
IABP transport issue
REPORT NO
076
1/4/21
Modified Ultrafiltration (MUF) Issue
REPORT NO
075
1/3/21
Electromagnetic interference with HLM
REPORT NO
074
1/2/21
HCU40 Failure
REPORT NO
073
1/2/21
HCU40 valve failures
REPORT NO
072
1/1/21
HCU40 failure
REPORT NO
071
1/1/21
LIVA NOVA STATEMENT ON CENTRIFUGAL HEAD
REPORT NO
070
1/1/21
Loose component on circuit
REPORT NO
069
1/12/20
Centrifugal Head noise
REPORT NO
068
1/11/20
Venous Reservoir disposable issue
REPORT NO
067
1/11/20
Failure of Heater Cooler due to leaking valves
REPORT NO
066
1/11/20
Failure of Valves on heater cooler unit
REPORT NO
065
1/9/20
Intra Aortic Balloon Pump (IABP) Failure
REPORT NO
064
1/9/20
Centrifugal head failure
REPORT NO
063
1/8/20
Centrifugal head issue 4
REPORT NO
062
1/8/20
Human Factors Issue
REPORT NO
061
1/8/20
LIVA NOVA RESPONSE TO REPORTED CENTRIFUGAL HEAD ISSUE
REPORT NO
060
1/8/20
Oxygenator Fibre Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.