PERFUSION REPORTS

REPORT
093
DATE
1/11/2021
TITLE
Detached bullet tip of sump-updated response Sep 2022
DETAILS OF INCIDENT

During mini-access MVR, surgeon became aware the bullet tip of the weighted sump had become detached and was in the LV cavity.

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

Liva Nova 20Fr weighted sump sucker "AtraSump"

Product code: SU-29602

Date of manufacture: July 2020

ACTION TAKEN

Surgeon removed bullet tip from ventricle with no issues. The sump was replaced without further problems and the faulty unit saved for inspection by Liva Nova.

RESPONSE FROM MANUFACTURER

Received response: Investigation summary

Liva Nova received a report stating that, during the procedure, the tip of the SU-29602 cannula came off. No defects were noted before starting the procedure, however the cannula detached during use but having recovered it immediately, no patient was affected.

A photo was provided as evidence of the defect, in particular the disconnected part was between: spring, inner tubing (PN 20742)and connector (PN 20741). The DHR review highlighted that the lot whose claimed product belong to was released as conform according to specifications. Up until April 2022, other similar complaints were recorded from different customers but on different cannula lots. The factors that could contribute to the reported issues are: (i) a failure during the manufacturing steps to apply the proximal connector on the tip assembly, performed by the operators, so that an adequate bonding was not guaranteed and / or (ii) a defective supplier component due to out of specs. The unit has been requested for investigation, and received in Arvada. Per MOP 6003-032, the connection site between the connector and the tubing should have a feature that did not appear to be present on returned unit, the seam between the tubing and the connector should have a visible bead of extra UV adhesive. While there is a thin layer of adhesive, it would appearthat this layer is flat and would hardly provide additional support to thisconnection. Moreover, the critical dimensions for this connection site were evaluated and the results revealed that the components used in this assembly appear to have been in specification. Therefore, no supplier issue occurred. Based on all facts mentioned, it cannot be ruled out that the root cause of th etip detachment is related to a manufacturing deviation during the assembly phase. The issue was addressed within CAPA-ARV-2022-0005. Further corrective action will be implemented through the CAPA process. Liva Nova maintains and document periodic customer events monitoring process in order to evaluate actions for products improvement.

ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
120
1/5/24
Disposable Incident
REPORT NO
119
Hardware: HCU Contamination
REPORT NO
118
12/9/23
Human Factors: ECMO Mobilisation
REPORT NO
117
27/9/23
Disposables: Oxygenator
REPORT NO
116
14/9/23
Disposable Issue: Reservoir in CPB circuit
REPORT NO
115
13/9/23
Disposable Error
REPORT NO
114
13/7/23
Dispoables
REPORT NO
112
20/2/23
HLM Hardware Error: S5 HLM - Dual roller cardioplegia
REPORT NO
111
10/2/23
Human Factors
REPORT NO
110
31/1/23
DISPOSABLE FAULT
REPORT NO
109
19/1/23
Disposable issue
REPORT NO
108
3/1/23
Disposable issue: Oxygenator leak
REPORT NO
107
6/12/22
Hardware fault. Electronic gas Blender failure
REPORT NO
106
1/11/22
ECG Interference
REPORT NO
105
9/9/22
Issue with use of disposable
REPORT NO
104
13/6/22
Human Factors
REPORT NO
103
25/7/22
Human Factors
REPORT NO
102
7/7/22
Disposable issue
REPORT NO
101
6/4/21
Irregular blood flow/foreign looking substance in Oxygenator
REPORT NO
100
20/4/22
Gas Line Filter Fault
REPORT NO
099
20/4/22
Human Factors: BiVAD Error
REPORT NO
098
4/4/22
REPORT NO
097
19/3/22
Disposables Issue
REPORT NO
096
18/3/22
Bubbles Sensor Activation
REPORT NO
095
1/1/22
Protamine Administration
REPORT NO
094
1/11/21
Oxygenation failure upon commencement of CPB
REPORT NO
093
1/11/21
Detached bullet tip of sump-updated response Sep 2022
REPORT NO
092
1/10/21
Vapourizer Issue
REPORT NO
091
1/10/21
Oyxgenator Leak
REPORT NO
090
1/10/21
Breach of chemical cleaner
Perfusion Report cards are shown in batches of 30 at any one time where available.
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.