Sorin Heat Exchanger Blood Water Leak
PROBLEM

The Sorin Group Italia have issued a Field Safety Corrective Action Notice concerning a number of complaints received from its customers regarding a leak between the blood and water compartments of the heat exchanger in a number of its oxygenators. The reported occurrence rate is very low, < 0.002%. Investigations suggest that the leaks may be caused by prolonged contact with sodium chloride (NaCl), causing micro holes to appear in the heat exchanger.

ACTION TAKEN

The Safety Committee would like to make the UK community aware of this situation  and suggest that the instructions be followed in the attached document.

DOWNLOADS
USEFUL LINKS

OTHER FSN REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view FSN Details
REPORT NO
017
1/8/2020
Getinge Device Recall IAB
REPORT NO
016
1/8/2020
Getinge HCU40 Replacment of Vacuum Valves
REPORT NO
015
1/7/2020
Getinge Replacement IFU's for HU35 and rotaflow console
REPORT NO
014
1/7/2020
Liva Nova Perceval Valve Accessory kit
REPORT NO
013
1/7/2020
Medtronic HeartWare HVAD system lubricant
REPORT NO
012
1/6/2020
Getinge Quadrox Holder FSN
REPORT NO
011
1/5/2020
Teleflex IABP
REPORT NO
010
1/4/2020
Medtronic HeartWare HVAD Systems
REPORT NO
009
1/3/2017
Infections Associated With HCUs In CPB & ECMO
REPORT NO
008
1/9/2015
Potential Mycobacterial Contamination Of HCUs
REPORT NO
007
1/2/2014
SWISSMEDIC: Water-Borne Infection In Implant Patients
REPORT NO
006
1/1/2014
Maquet Cardioplegia Temperature Reading Inaccuracies
REPORT NO
005
1/1/2012
Maquet Adhesive Failure On Oxygenator Blood Outlet Connector
REPORT NO
004
1/8/2011
Sorin Heat Exchanger Blood Water Leak
REPORT NO
003
1/6/2011
Maquet Transient Increase In Pump Speeds
REPORT NO
002
1/3/2011
Terumo CDI500 Possibility Of Parameter Drift
REPORT NO
001
1/2/2009
Datascope IABP Harness Not Fit For Purpose
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
019
018
017
016
015
014
013
012
011
010
009
008
007
006
005
004
003
002
001
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.