PERFUSION REPORTS

REPORT
012
DATE
1/2/2013
TITLE
Hypercoagulable Disorder May Have Led To Oxygenator Failure
DETAILS OF INCIDENT

A patient with complex medical history (IDDM and leg amputation 20 years previous) underwent CABG surgery.  The pre-operative TEG showed a relative hypercoagulable state.

Surgery proved complex and protracted due to poor targets and diffuse disease.  Three episodes of cardiopulmonary bypass (CPB) were necessary due to cardiovascular instability.

  • CPB run 1 was long but uneventful and was successfully terminated.  Protamine was administered along with platelets and FFP.  The CPB circuit was flushed with 1000ml Hartmann’s solution.  Patient instability led to re-instatement of CPB prior to which heparin was administered to the patient and a correct therapeutic ACT was obtained.  15KIU heparin was added to the pump prime and re-circulated.
  • CPB run 2, from the onset presented a failing oxygenator (low pO2 and low saturations) despite 100% oxygen.  The device was changed out successfully.  An extensive report from the manufacturer identified fibrin build up and no fault with the device.  Due to suspected heart failure, an IABP was inserted directly into the aorta (no femoral access possible) to assist weaning from CPB run 2.  Low flow was used for approximately 58 minutes to assist with IABP insertion and timing.  CPB was eventually terminated but cannulae were left in situ for 52 min.
  • CPB run 3 commenced due to continued instability.  The perfusionist was unable to achieve full flow and a hissing noise was heard. The arterial line filter was seen to have clotted off. CPB was terminated and the patient passed away.  Clots were seen in the arterial line and the aortic cannula.
DETAILS OF INJURY TO PATIENT
TYPE OF DEVICE/MANUFACTURER
ACTION TAKEN

An extensive report from the manufacturer identified fibrin build up within the first oxygenator and no fault with the device.   The second oxygenator and circuit were not kept for investigation.  As part of a tender process, the oxygenators in use for this case were new to the department.

A Comprehensive Investigation Report from the institution revealed that it has not been possible to conclusively identify contributory factors that led to the patient’s death. There was no inquest.  Two possible explanations in that report were:

  • Possible unidentified patient coagulopathy (? Antiphospholipid syndrome).
  • New oxygenator device in use during surgery.

It is important to note that neither explanation can be fully established or eliminated.

All standard protocol was followed during the procedure and all ACT results were well above the necessary threshold throughout the case. Inadvertent protamine administration was ruled out.

This incident has led a focus on awareness in the following areas:

  • A hypercoagulable disorder may play a role in oxygenator failure.  Any indication of such a state may call for improved point of care testing.   Anti-Phospholipid Syndrome which occurs in 2% of the population.  Antibodies present in patients with Antiphospholipid syndrome (APLS) are reported to affect a kaolin ACT.  These antibodies attack some fats and proteins in the blood causing it to clot.  Blood becomes stickier increasing the risk of blood clots.   In a patient with known APLS, a celite ACT is recommended with a higher target value and plotting of heparin/ACT titration curves pre-operatively to determine patient-specific target ACT values.  Heparin concentrations >3u/ml should be considered.  Use of the HepCon system would prove valuable.  Antifibrinolytic agents are not recommended.
  • When evaluating any new equipment, the whole surgical team should be made aware at the team briefing.
  • Any incidents, where faulty devices have been proved, should be reported to the MHRA as a precaution.
  • Any device or disposable with suspected fault should be retained for investigation along with all serial
    and lot numbers.
  • Consideration should be given to recirculation of the blood remaining in the pump during any significant period where CPB has been terminated but where the cannula may remain in situ.  This will prevent the blood from stagnating.  Consider taking a sample for ACT analysis and regularly flushing the cannulae.
  • During low flow CPB (used here for the assistance of IABP therapy), ensure that the minimum flow rate for the oxygenator in use is achieved or exceeded.  This minimum flow rate will be quoted in the manufacturers Instructions for Use and should be incorporated into the departmental protocols.

RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
120
1/5/24
Disposable Incident
REPORT NO
119
Hardware: HCU Contamination
REPORT NO
118
12/9/23
Human Factors: ECMO Mobilisation
REPORT NO
117
27/9/23
Disposables: Oxygenator
REPORT NO
116
14/9/23
Disposable Issue: Reservoir in CPB circuit
REPORT NO
115
13/9/23
Disposable Error
REPORT NO
114
13/7/23
Dispoables
REPORT NO
112
20/2/23
HLM Hardware Error: S5 HLM - Dual roller cardioplegia
REPORT NO
111
10/2/23
Human Factors
REPORT NO
110
31/1/23
DISPOSABLE FAULT
REPORT NO
109
19/1/23
Disposable issue
REPORT NO
108
3/1/23
Disposable issue: Oxygenator leak
REPORT NO
107
6/12/22
Hardware fault. Electronic gas Blender failure
REPORT NO
106
1/11/22
ECG Interference
REPORT NO
105
9/9/22
Issue with use of disposable
REPORT NO
104
13/6/22
Human Factors
REPORT NO
103
25/7/22
Human Factors
REPORT NO
102
7/7/22
Disposable issue
REPORT NO
101
6/4/21
Irregular blood flow/foreign looking substance in Oxygenator
REPORT NO
100
20/4/22
Gas Line Filter Fault
REPORT NO
099
20/4/22
Human Factors: BiVAD Error
REPORT NO
098
4/4/22
REPORT NO
097
19/3/22
Disposables Issue
REPORT NO
096
18/3/22
Bubbles Sensor Activation
REPORT NO
095
1/1/22
Protamine Administration
REPORT NO
094
1/11/21
Oxygenation failure upon commencement of CPB
REPORT NO
093
1/11/21
Detached bullet tip of sump-updated response Sep 2022
REPORT NO
092
1/10/21
Vapourizer Issue
REPORT NO
091
1/10/21
Oyxgenator Leak
REPORT NO
090
1/10/21
Breach of chemical cleaner
Perfusion Report cards are shown in batches of 30 at any one time where available.
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.