DETAILS OF INCIDENT

All pre-bypass checks complete. Bypass initiated, cross clamp applied and retrograde cardioplegia administration started. Unable to get sufficient flow rate due to coronary sinus pressure therefore stopped and moved to attempt antegrade cardioplegia via aortic root. When cardioplegia started the S5 console alarmed and displayed 'Pump malfunction' and it was noted that the crystalloid pump of the dual pump cardioplegia pump was not spinning.

Surgeon immediately informed that there was a serious failure and we were unable to deliver cardioplegia.

DETAILS OF INJURY TO PATIENT
TYPE OF DEVICE/MANUFACTURER

S5 Double Roller Pump85 / SN10E53864/ Date of Manufacture 2016-02-22

ACTION TAKEN

Surgeon immediately informed that there was a serious failure and we were unable to deliver cardioplegia. Cross clamp was removed and the heart was re-perfused. The cardioplegia pump was power cycled in isolation, testing was carried out and found that the issue had resolved. The decision was made to continue with the operation, which otherwise was uneventful.

RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION

The safety committee has reviewed this incident and acknowledge that the older S5 pumps may have a trend with software errors that can account for these issues. Age and life span of the pump is something that must be monitored along with the incident rate and problems.

However, the pump was acquired in Feb 2016, so 7 years old.There have been no further issues with the pump. All pumps are regularlyserviced and maintained.

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
120
1/5/24
Disposable Incident
REPORT NO
119
Hardware: HCU Contamination
REPORT NO
118
12/9/23
Human Factors: ECMO Mobilisation
REPORT NO
117
27/9/23
Disposables: Oxygenator
REPORT NO
116
14/9/23
Disposable Issue: Reservoir in CPB circuit
REPORT NO
115
13/9/23
Disposable Error
REPORT NO
114
13/7/23
Dispoables
REPORT NO
112
20/2/23
HLM Hardware Error: S5 HLM - Dual roller cardioplegia
REPORT NO
111
10/2/23
Human Factors
REPORT NO
110
31/1/23
DISPOSABLE FAULT
REPORT NO
109
19/1/23
Disposable issue
REPORT NO
108
3/1/23
Disposable issue: Oxygenator leak
REPORT NO
107
6/12/22
Hardware fault. Electronic gas Blender failure
REPORT NO
106
1/11/22
ECG Interference
REPORT NO
105
9/9/22
Issue with use of disposable
REPORT NO
104
13/6/22
Human Factors
REPORT NO
103
25/7/22
Human Factors
REPORT NO
102
7/7/22
Disposable issue
REPORT NO
101
6/4/21
Irregular blood flow/foreign looking substance in Oxygenator
REPORT NO
100
20/4/22
Gas Line Filter Fault
REPORT NO
099
20/4/22
Human Factors: BiVAD Error
REPORT NO
098
4/4/22
REPORT NO
097
19/3/22
Disposables Issue
REPORT NO
096
18/3/22
Bubbles Sensor Activation
REPORT NO
095
1/1/22
Protamine Administration
REPORT NO
094
1/11/21
Oxygenation failure upon commencement of CPB
REPORT NO
093
1/11/21
Detached bullet tip of sump-updated response Sep 2022
REPORT NO
092
1/10/21
Vapourizer Issue
REPORT NO
091
1/10/21
Oyxgenator Leak
REPORT NO
090
1/10/21
Breach of chemical cleaner
Perfusion Report cards are shown in batches of 30 at any one time where available.
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.