PERFUSION REPORTS

REPORT
108
DATE
3/1/2023
TITLE
Disposable issue: Oxygenator leak
DETAILS OF INCIDENT

During standard priming checks there was a leak noted from the housing of the oxygenator. The heater cooler was emptied to ensure there wasn't a leak from the heater cooler. The oxygenator was pressurised with recirculation lines clamped and the inlet to the oxygenator also clamped (to isolate the pump boot) and the leak persisted with a constant pressure drop noted in the oxygenator.

DETAILS OF INJURY TO PATIENT

No injury.

TYPE OF DEVICE/MANUFACTURER

Liva Nova Oxygenator product code: IN01108H

ACTION TAKEN

A replacement set was utilised and the faulty set was put aside for analysis by the manufacturer, manufacturer was informed and complaint started. Ulysses (datix) completed for trust tracking.

RESPONSE FROM MANUFACTURER

Returned part (SN 592786-0092) was visually inspected and no damaged component was identified.

A subsequent dye-leak test on blood side was carried out to verify the integrity of the device.

The unit was filled with methylene blue and pressurized by sealing all the outlet ports, in line with the maximum internal pressure value permitted in blood path (1 Bar) as per product specifications.

The internal pressure value was kept monitored with a pressure gauge connected to the circuit at the level of the inlet.

In the early filling phase, the inner core of oxygenator fiber bundle was found to be broken (it turned blue) and spillage of dye solution from open slot region of the device was reproduced.

Based on observed findings, reported failure was confirmed and traced back to damaged fiber/s of the oxygenator module.

In such a case, fluid (priming or blood) enters into the ruptured capillary and leaks out of gas escape compartment following gas flow direction.

Considering that 100% of the units in production line is subjected to final leak test and that involved unit successfully passed the in-process check and no trend has been registered for similar cases, it cannot be excluded that fiber bundle capillary manufacturing variability with contribution of sterilization thermal and/or transportation mechanical stress could have led to breakage during use.

No specific action is currently deemed as necessary.

LivaNova maintains and documents periodic customer events monitoring process in order to evaluate actions for products improvement.

ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
120
1/5/24
Disposable Incident
REPORT NO
119
Hardware: HCU Contamination
REPORT NO
118
12/9/23
Human Factors: ECMO Mobilisation
REPORT NO
117
27/9/23
Disposables: Oxygenator
REPORT NO
116
14/9/23
Disposable Issue: Reservoir in CPB circuit
REPORT NO
115
13/9/23
Disposable Error
REPORT NO
114
13/7/23
Dispoables
REPORT NO
112
20/2/23
HLM Hardware Error: S5 HLM - Dual roller cardioplegia
REPORT NO
111
10/2/23
Human Factors
REPORT NO
110
31/1/23
DISPOSABLE FAULT
REPORT NO
109
19/1/23
Disposable issue
REPORT NO
108
3/1/23
Disposable issue: Oxygenator leak
REPORT NO
107
6/12/22
Hardware fault. Electronic gas Blender failure
REPORT NO
106
1/11/22
ECG Interference
REPORT NO
105
9/9/22
Issue with use of disposable
REPORT NO
104
13/6/22
Human Factors
REPORT NO
103
25/7/22
Human Factors
REPORT NO
102
7/7/22
Disposable issue
REPORT NO
101
6/4/21
Irregular blood flow/foreign looking substance in Oxygenator
REPORT NO
100
20/4/22
Gas Line Filter Fault
REPORT NO
099
20/4/22
Human Factors: BiVAD Error
REPORT NO
098
4/4/22
REPORT NO
097
19/3/22
Disposables Issue
REPORT NO
096
18/3/22
Bubbles Sensor Activation
REPORT NO
095
1/1/22
Protamine Administration
REPORT NO
094
1/11/21
Oxygenation failure upon commencement of CPB
REPORT NO
093
1/11/21
Detached bullet tip of sump-updated response Sep 2022
REPORT NO
092
1/10/21
Vapourizer Issue
REPORT NO
091
1/10/21
Oyxgenator Leak
REPORT NO
090
1/10/21
Breach of chemical cleaner
Perfusion Report cards are shown in batches of 30 at any one time where available.
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.