The red sucker with one-way valve was used as a vent in CABG. Red vent turned off and cardioplegia given. Approximately 2 minutes after this, with vent still off, blood started quickly dripping out of one-way vent.
No injury occurred to patient or staff.
Eurosets EU82647 Adult tubing set. LOT/Batch 012016506. Manufacture date 2025-09-26, Expiry date 2028-09-25.
Clamped out and cut away leaking valve, replaced with 1/4" connector and green needle with line to vent. Kept defected valve, flushed it with water and contacted manufacturer. Sent back to Eurosets.