PERFUSION REPORTS

REPORT
144
DATE
29/7/2026
TITLE
Oxygen sensor
DETAILS OF INCIDENT

The sensor had a leak at its connection site. Whilst it still read the called for percentage of oxygen, it did not identify the significalnt reduction in flow. This was extremely misleading as it's use in identifying a cause for extreme hypercapnia resulted in suspicion of oxygenator failure. As a side note this happened 3 times before the root cause was identified.

DETAILS OF INJURY TO PATIENT
TYPE OF DEVICE/MANUFACTURER

Honeywell/Envitec Oxygen Sensor

ACTION TAKEN
RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION

Example of how a piece of safety equipment can reduce the safety of the procedure.

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
144
29/7/26
Oxygen sensor
REPORT NO
143
1/6/26
Case report - runaway pressure module
REPORT NO
142
6/7/26
Isoflurane vaporiser
REPORT NO
141
26/6/26
EOPA cannula
REPORT NO
140
12/6/26
Isoflurane vaporiser
REPORT NO
139
20/5/26
Faulty pressure relief valve
REPORT NO
138
20/4/26
Oxygenator Changeout
REPORT NO
137
12/11/25
Disposables: vent valves
REPORT NO
136
28/1/26
Hardware: ACT machine
REPORT NO
135
3/2/26
Human Factors
REPORT NO
134
17/12/25
Disposables: Faulty vent valve
REPORT NO
133
7/7/25
Disposables: Oxygenator Failure
REPORT NO
132
11/6/25
Hardware: Maquet cardiosave hybrid
REPORT NO
131
23/10/24
Human Factors
REPORT NO
130
19/5/25
Hardware: Level Alarm
REPORT NO
129
3/5/25
Human Factors
REPORT NO
128
25/5/25
Single Station Selectatec Type Backbar for Vapouriser. 9000-001
REPORT NO
127
4/1/25
Eurosets Oxygenator leak
REPORT NO
126
1/3/25
Oxygenator Failure
REPORT NO
125
11/2/25
Hardware Error: Bubble Detector
REPORT NO
124
1/10/24
HCU/ Hardware incident/ Human Factors
REPORT NO
123
5/7/24
LivaNova S5 Roller Pump Fault
REPORT NO
122
5/4/24
Oxygenator Failure
REPORT NO
121
8/3/24
Disposables: Eurosets Adult ECMO Oxygenator
REPORT NO
120
1/5/24
Oxygenator Failure
REPORT NO
119
Hardware: HCU Contamination
REPORT NO
118
12/9/23
Human Factors: ECMO Mobilisation
REPORT NO
117
27/9/23
Disposables: Oxygenator failure
REPORT NO
116
14/9/23
Disposable Issue: Reservoir in CPB circuit
REPORT NO
115
13/9/23
Disposable Error/ Eurosets New Born ECMO oxygenator (EU5056) leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
144
143
142
141
140
139
138
137
136
135
134
133
132
131
130
129
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.