PERFUSION REPORTS

REPORT
081
DATE
1/4/2021
TITLE
S5 Double Roller Pump incident
DETAILS OF INCIDENT

During administration of the first dose of cardioplegia the pump that mixes the cardioplegia drug with blood stopped. This was observed by the perfusionist running the heart lung machine. The blood pump (master) to which the cardioplegia pump is slaved continued to run.

The control knob of the master was turned to off and then twisted to restart administration but the cardioplegia pump would not start.

The surgical team was alerted and the settings of the cardioplegia pumps and module reviewed but with no observable cause.

The cardioplegia dual pump base was turned off and on again and the pumps restarted. This time the cardioplegia pump restarted but the flow was much faster than the flow ratio set in the cardioplegia module.

The perfusionist administered cardioplegia intermittently due to the high flow to achieve cardiac arrest. After an appropriate dose was administered, the perfusionist sought a second opinion of a perfusion colleague.

Within the settings on the cardioplegia pump A the master slave ratio was displayed as "NaN". By changing the flow ratio on the cardioplegia module this value was changed to a corresponding percentage and the normal operation returned. The pump was be quarantined and reported to the supplier and the MHRA.

DETAILS OF INJURY TO PATIENT
TYPE OF DEVICE/MANUFACTURER

Liva Nova S5 Heart Lung Machine Double Roller Pump

ACTION TAKEN

As described above

RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
115
13/9/23
Disposable Error
REPORT NO
114
13/7/23
Dispoables
REPORT NO
111
10/2/23
Human Factors
REPORT NO
110
31/1/23
DISPOSABLE FAULT
REPORT NO
109
19/1/23
Disposable issue
REPORT NO
108
3/1/23
Disposable issue: Oxygenator leak
REPORT NO
107
6/12/22
Hardware fault. Electronic gas Blender failure
REPORT NO
106
1/11/22
ECG Interference
REPORT NO
105
9/9/22
Issue with use of disposable
REPORT NO
104
13/6/22
Human Factors
REPORT NO
103
25/7/22
Human Factors
REPORT NO
102
7/7/22
Disposable issue
REPORT NO
101
6/4/21
Irregular blood flow/foreign looking substance in Oxygenator
REPORT NO
100
20/4/22
Gas Line Filter Fault
REPORT NO
099
20/4/22
Human Factors: BiVAD Error
REPORT NO
098
4/4/22
REPORT NO
097
19/3/22
Disposables Issue
REPORT NO
096
18/3/22
Bubbles Sensor Activation
REPORT NO
095
1/1/22
Protamine Administration
REPORT NO
094
1/11/21
Oxygenation failure upon commencement of CPB
REPORT NO
093
1/11/21
Detached bullet tip of sump-updated response Sep 2022
REPORT NO
092
1/10/21
Vapourizer Issue
REPORT NO
091
1/10/21
Oyxgenator Leak
REPORT NO
090
1/10/21
Breach of chemical cleaner
REPORT NO
089
1/9/21
Sechrist Gas Blender Sweep on ECMO
REPORT NO
088
1/9/21
Oxygenator High Pressure Events
REPORT NO
087
1/10/21
Leaking one way valve
REPORT NO
086
1/7/21
Electronic Gas Blender Fault
REPORT NO
085
1/2/21
Protamine Administration Issue
REPORT NO
084
1/6/21
Damaged roller pump lid
Perfusion Report cards are shown in batches of 30 at any one time where available.
115
114
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
019
018
017
016
015
014
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.