REPORT
072
DATE
June 2021
TITLE
Damaged roller pump lid
DETAILS OF INCIDENT

The roller pump lid hinge had cracked and the piece dropped into the raceway. The piece was sharp enough to potentially puncture the silicone tubing. The broken piece wasn't noticed during set up but it was spotted whilst priming.

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

Liva Nova S5

ACTION TAKEN

Good catch: Piece removed-highlights importance of regular checking hardware for wear and tear and replacing damaged components to prevent any incidents or malfunction.

RESPONSE FROM MANUFACTURER

n/a

ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO

see photos attached

SUPPORTING DOCUMENT

OTHER PERFUSION REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
077
DATE
Sep 2021
Sechrist Gas Blender Sweep on ECMO
REPORT NO
076
DATE
Sep 2021
Oxygenator High Pressure Events
REPORT NO
075
DATE
Oct 2021
Leaking one way valve
REPORT NO
074
DATE
July 2021
Electronic Gas Blender Fault
REPORT NO
073
DATE
February 2021
Protamine Administration Issue
REPORT NO
072
DATE
June 2021
Damaged roller pump lid
REPORT NO
071
DATE
June 2021
IABP malfunction
REPORT NO
070
DATE
Jan 2021
Accidental cardioplegia administration
REPORT NO
069
DATE
June 2021
Sampling manifold valve fault
REPORT NO
068
DATE
May 2021
Human factors issue
REPORT NO
067
DATE
June 2021
Centrimag battery module failure
REPORT NO
066
DATE
April 2021
S5 Double Roller Pump incident
BACK TO SAFETY ARCHIVE