REPORT
041
DATE
Feb 2020
TITLE
Arterial Line Filter Leak
DETAILS OF INCIDENT

After priming the circuit with a blood prime prior to CPB, a drip of blood was noticed on the floor. Upon closer inspection, the arterial line filter had leaked all around the seal.

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

Liva Nova Arterial Line Filter

Product Code: C23500

Lot/Batch: 1910240140

ACTION TAKEN

A new tubing set was opened, the filter was isolated from the circuit and removed and a new arterial line filter was inserted. The original faulty filter was drained of blood and stored according to department protocol for the manufacturer to collect. The manufacturer was contacted and an electronic incident form was completed complying with trust guidelines.

RESPONSE FROM MANUFACTURER

See Attached

ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO

See attached

SUPPORTING DOCUMENT

OTHER PERFUSION REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
049
DATE
August 2020
LIVA NOVA RESPONSE TO REPORTED CENTRIFUGAL HEAD ISSUE
REPORT NO
048
DATE
August 2020
Oxygenator Fibre Leak
REPORT NO
047
DATE
May 2020
Failure to remove CO2 effectively
REPORT NO
046
DATE
July 2020
Centrifugal Head Issue 3
REPORT NO
045
DATE
June 2020
Centrifugal Head Issue 2
REPORT NO
044
DATE
Jan 2020
Centrifugal Head Issue
REPORT NO
043
DATE
Feb 2020
Faulty Pressure Relief Valves
REPORT NO
042
DATE
Apr 2020
Leaking ALF
REPORT NO
041
DATE
Feb 2020
Arterial Line Filter Leak
REPORT NO
040
DATE
Mar 2019
Gas Line Filter Issue
REPORT NO
039
DATE
Dec 2019
Dual Lumen Cannula Movement
REPORT NO
038
DATE
Oct 2019
Venous Cannula Damaged
BACK TO SAFETY ARCHIVE