PERFUSION REPORTS

REPORT
016
DATE
TITLE
Additional Sucker Incorrectly Placed In Raceway
DETAILS OF INCIDENT

Trainee Perfusionist added an extra sucker to a pre built circuit & failed to complete the 3 perfusion sucker checks. The 4th and final sucker check, a visual wet test of the suckers with the scrub nurse was only completed for 2 of the 3 suckers i.e. not the additional sucker.

The additional sucker was used as a vent and after x clamping the surgeon asked for the vent to be turned on and noted it was blowing not sucking, this was switched off and corrected. The surgeon was adamant that the Aorta was X clamped before the vent was switched on and that no air entered the patient.

DETAILS OF INJURY TO PATIENT

Post operatively the patient showed Lt sided stroke and the family was informed of the incident involving the reversed suction tubing.

The patient and family have been kept fully informed as to the possible reasons for the stroke - and that stroke is a known risk from cardiac surgery and cardiopulmonary bypass.

The family has also been kept fully informed of the mistakes made during setup and checking of the suckers.

TYPE OF DEVICE/MANUFACTURER

N/A

ACTION TAKEN

A Hospital Incident Report was filed on the day of the incident. A panel met and discussed the incident and decided to conduct a full SUI investigation with outcome of investigation sent to family and the SHA.

A full review of practices has been undertaken, this has resulted in a number of changes to procedure and/or protocols.

RESPONSE FROM MANUFACTURER

N/A

ADDTIONAL INFORMATION

N/A

SUPPORTING VISUALS/VIDEO

N/A

SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
059
1/5/20
Failure to remove CO2 effectively
REPORT NO
058
1/7/20
Centrifugal Head Issue 3
REPORT NO
057
1/6/20
Centrifugal Head Issue 2
REPORT NO
056
1/1/20
Centrifugal Head Issue
REPORT NO
055
1/2/20
Faulty Pressure Relief Valves
REPORT NO
054
1/4/20
Leaking ALF
REPORT NO
053
1/2/20
Arterial Line Filter Leak
REPORT NO
052
1/3/19
Gas Line Filter Issue
REPORT NO
051
1/12/19
Dual Lumen Cannula Movement
REPORT NO
050
1/10/19
Venous Cannula Damaged
REPORT NO
049
1/12/19
Oxygenation Issue
REPORT NO
048
1/8/19
Roller Pump Failure
REPORT NO
047
1/12/19
Broken Cannula Dilator
REPORT NO
046
6/12/19
Gas Blender Error
REPORT NO
045
20/11/19
Tubing Detachments
REPORT NO
044
27/11/19
Interruption to IABP counterpulsation
REPORT NO
043
1/7/19
Getinge Mal-Position Of IABP Catheter
REPORT NO
042
1/2/19
Avalon Dual Lumen Cannula Crack
REPORT NO
041
1/2/19
Custodial Rubber Fragmentation In Spiked Bag
REPORT NO
040
1/10/18
Reservoir Clot
REPORT NO
039
1/10/18
Pall Clot In Cardioplegia Circuit
REPORT NO
038
1/10/18
Maquet HL20 Cardioplegia Pump Error
REPORT NO
037
1/10/17
Temed D200 Haemoconcentrator Clot
REPORT NO
036
1/10/17
Oxygenator issue
REPORT NO
035
1/10/17
Medtronic Fusion Oxygenator Blood To Gas Leak
REPORT NO
034
Possible Medtronic Fusion Oxygenator Failure
REPORT NO
033
Fannin Heparin Ampules Shattering
REPORT NO
032
Sucker Placed In Raceway In Wrong Direction
REPORT NO
031
High Transmembrane Pressure Gradients
REPORT NO
030
Faulty Connector On ECMO Soft Shell Reservoir Bag
Perfusion Report cards are shown in batches of 30 at any one time where available.
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.