PERFUSION REPORTS

REPORT
029
DATE
TITLE
Sorin Inspire 8F Blood Leak From Temperature Port
DETAILS OF INCIDENT

The oxygenator (a Sorin Inspire 8F) was setup as per our normal and pressure tested for 10 minutes prior to priming. All routine pressure / occlusion checks were performed and passed. The pump was left to recirculate during mammary harvest at 140mmHg. Upon return, we noticed several micro bubbles in the oxygenator near the white recirculation port but no visible leak which was de-aired with flow and pressure being applied. A very small leak was noted at this point but the location of the leak was difficult to visualize. Therefore, all luer lock connectors for the sampling manifold and both red and white recirculation ports were tightened, at which point the leak seemed to stop.



We continued with an arterial R.A.P which was uneventful. The venous tubing was then R.A.P.ed and it was this at this point where a fast and almost continuous blood leak was noted out the bottom of the oxygenator seemingly originating from inside or around the temperature port.


DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

N/A

ACTION TAKEN

Lines were clamped off to patient and oxygenator was changed out.


RESPONSE FROM MANUFACTURER

N/A

ADDTIONAL INFORMATION

N/A

SUPPORTING VISUALS/VIDEO

N/A

SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
151
15/9/26
Gas line disconnection
REPORT NO
150
21/10/26
Governance issue of undocumented IABP therapy
REPORT NO
149
15/9/26
Oxygen sensor gas leak
REPORT NO
148
14/9/26
Consumables - cracked cardioplegia table line
REPORT NO
147
9/9/26
Isoflurane vaporiser
REPORT NO
146
19/5/26
Oxygen sensor gas leak
REPORT NO
145
6/8/26
Disposables: Faulty cardioplegia cannula
REPORT NO
144
29/7/26
Oxygen sensor
REPORT NO
143
1/6/26
Case report - runaway pressure module
REPORT NO
142
6/7/26
Isoflurane vaporiser
REPORT NO
141
26/6/26
EOPA cannula
REPORT NO
140
12/6/26
Isoflurane vaporiser
REPORT NO
139
20/5/26
Faulty pressure relief valve
REPORT NO
138
20/4/26
Oxygenator Changeout
REPORT NO
137
12/11/25
Disposables: vent valves
REPORT NO
136
28/1/26
Hardware: ACT machine
REPORT NO
135
3/2/26
Human Factors
REPORT NO
134
17/12/25
Disposables: Faulty vent valve
REPORT NO
133
7/7/25
Disposables: Oxygenator Failure
REPORT NO
132
11/6/25
Hardware: Maquet cardiosave hybrid
REPORT NO
131
23/10/24
Human Factors
REPORT NO
130
19/5/25
Hardware: Level Alarm
REPORT NO
129
3/5/25
Human Factors
REPORT NO
128
25/5/25
Single Station Selectatec Type Backbar for Vapouriser. 9000-001
REPORT NO
127
4/1/25
Eurosets Oxygenator leak
REPORT NO
126
1/3/25
Oxygenator Failure
REPORT NO
125
11/2/25
Hardware Error: Bubble Detector
REPORT NO
124
1/10/24
HCU/ Hardware incident/ Human Factors
REPORT NO
123
5/7/24
LivaNova S5 Roller Pump Fault
REPORT NO
122
5/4/24
Oxygenator Failure
Perfusion Report cards are shown in batches of 30 at any one time where available.
151
150
149
148
147
146
145
144
143
142
141
140
139
138
137
136
135
134
133
132
131
130
129
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.