Vent Pump Boot Tubing Inserted In Wrong Way Round
DETAILS OF INCIDENT
  • Vent Pump Boot tubing inserted in pump the wrong way round. Wet test failed to show up.
  • New bypass machines were in use.
  • The tubing pack was still under modification to fit new machines. Vent sucker required extending to reach new reservoir position.
  • Wet test performed by scrub person and confirmed as being okay. PA vent inserted, surgeon immediately reports that there was no suction. Suction was stopped and the occlusion was checked, but still no suction. Wet test repeated and the vent observed blowing bubbles. The perfusionist spots a pump boot error and reverses the boot into the correct orientation. Wet test repeated, vent observed sucking. Case proceeds as normal. Anaesthetist was viewing TOE throughout and reported no air. Patient recovered as normal.
DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

N/A

ACTION TAKEN
  • Discussed and raised awareness of potential errors with the introduction of new equipment. Modification of tubing sets to eliminate the need to extend the vent sucker and also increase the visibility of the connector for the vent at the inlet side of the pump. Review of all sucker tubing indicators in the operative field.
  • Increased robustness of wet test.
  • It was felt that the failure of the wet test may have been due to the use of a connecting silicone 'U' bend that connects 3/8th inch arterial to 1/2 inch venous in the tubing sash. This is cut out of the system by the scrub person and retains some priming fluid which was being dipped into for the wet test. It may have indicated a positive result as a sucker tip pushed into the 3/8th inch side of the 'U' could displace fluid up the 1/2 inch side which would then naturally swing back up the sucker tip. Based on this the wet test is now performed from an open jug and visualised and confirmed by the perfusionist and the scrub person.
  • One way valves in all suction lines was highlighted as well as a final dip test by the operator prior to insertion of intra cardiac suction. Neither of these measures are currently in place.
  • Better indicators on the pump for direction off flow are being explored.
RESPONSE FROM MANUFACTURER

N/A

ADDTIONAL INFORMATION

N/A

SUPPORTING VISUALS/VIDEO

N/A

SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
059
1/5/20
Failure to remove CO2 effectively
REPORT NO
058
1/7/20
Centrifugal Head Issue 3
REPORT NO
057
1/6/20
Centrifugal Head Issue 2
REPORT NO
056
1/1/20
Centrifugal Head Issue
REPORT NO
055
1/2/20
Faulty Pressure Relief Valves
REPORT NO
054
1/4/20
Leaking ALF
REPORT NO
053
1/2/20
Arterial Line Filter Leak
REPORT NO
052
1/3/19
Gas Line Filter Issue
REPORT NO
051
1/12/19
Dual Lumen Cannula Movement
REPORT NO
050
1/10/19
Venous Cannula Damaged
REPORT NO
049
1/12/19
Oxygenation Issue
REPORT NO
048
1/8/19
Roller Pump Failure
REPORT NO
047
1/12/19
Broken Cannula Dilator
REPORT NO
046
6/12/19
Gas Blender Error
REPORT NO
045
20/11/19
Tubing Detachments
REPORT NO
044
27/11/19
Interruption to IABP counterpulsation
REPORT NO
043
1/7/19
Getinge Mal-Position Of IABP Catheter
REPORT NO
042
1/2/19
Avalon Dual Lumen Cannula Crack
REPORT NO
041
1/2/19
Custodial Rubber Fragmentation In Spiked Bag
REPORT NO
040
1/10/18
Reservoir Clot
REPORT NO
039
1/10/18
Pall Clot In Cardioplegia Circuit
REPORT NO
038
1/10/18
Maquet HL20 Cardioplegia Pump Error
REPORT NO
037
1/10/17
Temed D200 Haemoconcentrator Clot
REPORT NO
036
1/10/17
Oxygenator issue
REPORT NO
035
1/10/17
Medtronic Fusion Oxygenator Blood To Gas Leak
REPORT NO
034
Possible Medtronic Fusion Oxygenator Failure
REPORT NO
033
Fannin Heparin Ampules Shattering
REPORT NO
032
Sucker Placed In Raceway In Wrong Direction
REPORT NO
031
High Transmembrane Pressure Gradients
REPORT NO
030
Faulty Connector On ECMO Soft Shell Reservoir Bag
Perfusion Report cards are shown in batches of 30 at any one time where available.
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.