PERFUSION REPORTS

REPORT
044
DATE
27/11/2019
TITLE
Interruption to IABP counterpulsation
DETAILS OF INCIDENT

There was an interruption to IABP counter pulsation therapy out of hours. On arrival, alarm message was displayed "failure to pump". The gas line was inspected with no obstruction. Several attempts at re-institution of therapy were made unsuccessfully. A change of console was carried out and therapy was recommenced.

DETAILS OF INJURY TO PATIENT

Interruption to counter pulsation with IABP in standby for approximately 50 minutes.

TYPE OF DEVICE/MANUFACTURER

Getinge Cardiosave Hybrid

ACTION TAKEN

On arrival of the on-call Clinical Perfusion Scientist, the console was changed out and therapy was recommenced. It was not deemed appropriate/necessary to change IAB catheter.

The affected console was taken out of service and repaired. It was noted post service that it was a failure of the safety disk.

RESPONSE FROM MANUFACTURER

ADDTIONAL INFORMATION

Upon reflection what do you think went well?

The on-call Clinical Perfusion Scientist arrived promptly (approx 30 mins out of hours) and the console was changed out quickly with no issues.

New action plans/changes as a result of the event?

A teaching session with senior intensive care nursing staff (zone leaders) on troubleshooting and demystifying the IABP console to be arranged.

Troubleshooting flow charts have been added to the consoles to try minimise any downtime in patient therapy, should interruption occur. It is intended to have annual refreshers for this group of staff.

MHRA Informed

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
083
1/6/21
IABP malfunction
REPORT NO
082
1/1/21
Accidental cardioplegia administration
REPORT NO
081
1/4/21
S5 Double Roller Pump incident
REPORT NO
080
1/5/21
Human factors issue
REPORT NO
079
1/6/21
Centrimag battery module failure
REPORT NO
078
1/6/21
Sampling manifold valve fault
REPORT NO
077
1/5/21
IABP transport issue
REPORT NO
076
1/4/21
Modified Ultrafiltration (MUF) Issue
REPORT NO
075
1/3/21
Electromagnetic interference with HLM
REPORT NO
074
1/2/21
HCU40 Failure
REPORT NO
073
1/2/21
HCU40 valve failures
REPORT NO
072
1/1/21
HCU40 failure
REPORT NO
071
1/1/21
LIVA NOVA STATEMENT ON CENTRIFUGAL HEAD
REPORT NO
070
1/1/21
Loose component on circuit
REPORT NO
069
1/12/20
Centrifugal Head noise
REPORT NO
068
1/11/20
Venous Reservoir disposable issue
REPORT NO
067
1/11/20
Failure of Heater Cooler due to leaking valves
REPORT NO
066
1/11/20
Failure of Valves on heater cooler unit
REPORT NO
065
1/9/20
Intra Aortic Balloon Pump (IABP) Failure
REPORT NO
064
1/9/20
Centrifugal head failure
REPORT NO
063
1/8/20
Centrifugal head issue 4
REPORT NO
062
1/8/20
Human Factors Issue
REPORT NO
061
1/8/20
LIVA NOVA RESPONSE TO REPORTED CENTRIFUGAL HEAD ISSUE
REPORT NO
060
1/8/20
Oxygenator Fibre Leak
REPORT NO
059
1/5/20
Failure to remove CO2 effectively
REPORT NO
058
1/7/20
Centrifugal Head Issue 3
REPORT NO
057
1/6/20
Centrifugal Head Issue 2
REPORT NO
056
1/1/20
Centrifugal Head Issue
REPORT NO
055
1/2/20
Faulty Pressure Relief Valves
REPORT NO
054
1/4/20
Leaking ALF
Perfusion Report cards are shown in batches of 30 at any one time where available.
115
114
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
019
018
017
016
015
014
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.