PERFUSION REPORTS

REPORT
067
DATE
1/11/2020
TITLE
Failure of Heater Cooler due to leaking valves
DETAILS OF INCIDENT

I received a call out of hours from a member of theatre staff, the HCU40 in cardiac theatres was alarming "power failure" and sparks were seen coming from the bottom of the device. I advised the nurse to switch off the HCU.

There had been 2 Field Safety Notices issued in recent months recommending replacement of valves and crimping of power cables, however the recommendation in the safety notice had been to continue use as normal.

(Please see picture attached in additional information)

(Please see FSN 028)

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

Getinge HCU40

ACTION TAKEN

The next day the unit was removed from service, the engineer was informed and a repair/replacment was arranged.

The unit valves were replaced and the unit was put back into service.

RESPONSE FROM MANUFACTURER

n/a

ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
089
1/9/21
Sechrist Gas Blender Sweep on ECMO
REPORT NO
088
1/9/21
Oxygenator High Pressure Events
REPORT NO
087
1/10/21
Leaking one way valve
REPORT NO
086
1/7/21
Electronic Gas Blender Fault
REPORT NO
085
1/2/21
Protamine Administration Issue
REPORT NO
084
1/6/21
Damaged roller pump lid
REPORT NO
083
1/6/21
IABP malfunction
REPORT NO
082
1/1/21
Accidental cardioplegia administration
REPORT NO
081
1/4/21
S5 Double Roller Pump incident
REPORT NO
080
1/5/21
Human factors issue
REPORT NO
079
1/6/21
Centrimag battery module failure
REPORT NO
078
1/6/21
Sampling manifold valve fault
REPORT NO
077
1/5/21
IABP transport issue
REPORT NO
076
1/4/21
Modified Ultrafiltration (MUF) Issue
REPORT NO
075
1/3/21
Electromagnetic interference with HLM
REPORT NO
074
1/2/21
HCU40 Failure
REPORT NO
073
1/2/21
HCU40 valve failures
REPORT NO
072
1/1/21
HCU40 failure
REPORT NO
071
1/1/21
LIVA NOVA STATEMENT ON CENTRIFUGAL HEAD
REPORT NO
070
1/1/21
Loose component on circuit
REPORT NO
069
1/12/20
Centrifugal Head noise
REPORT NO
068
1/11/20
Venous Reservoir disposable issue
REPORT NO
067
1/11/20
Failure of Heater Cooler due to leaking valves
REPORT NO
066
1/11/20
Failure of Valves on heater cooler unit
REPORT NO
065
1/9/20
Intra Aortic Balloon Pump (IABP) Failure
REPORT NO
064
1/9/20
Centrifugal head failure
REPORT NO
063
1/8/20
Centrifugal head issue 4
REPORT NO
062
1/8/20
Human Factors Issue
REPORT NO
061
1/8/20
LIVA NOVA RESPONSE TO REPORTED CENTRIFUGAL HEAD ISSUE
REPORT NO
060
1/8/20
Oxygenator Fibre Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.