PERFUSION REPORTS

REPORT
109
DATE
19/1/2023
TITLE
Disposable issue
DETAILS OF INCIDENT

While priming the Eurosets Adult A.L. one ECMO oxygenator (EU5019) it was noted that there was a leak from the bottom of the oxygenator. the leak appeared to originate form the seam between the fibres and the plastic housing, suspected lack of bonding. The Oxygenator was changed out during priming. A representative from the company was contacted and an inspection completed.

Eurosets testing:

We set-up the circuit: the reservoir is filled with 500 ml of physiological water; the oxygenator module is filled form the reservoir using a peristaltic pump (Q = 6 l/min) and remained in loop for 10 minutes. We notice a slow dripping in the lower part of the oxygenator, in front of the arterial blood out. To identify the exact point of leakage, the device was sectioned removing the Gas-escape lid, refilled with water, and then pressurized.

The point of loss occurred due to the fiber breakage of the last winding at the blood outlet.

The leak reported during the priming phase can be related to a fiber rupture occurred after Eurosets manufacturing and test phases. however, further improvement processes have been implemented to reduce this occurrence.

The safety committee have reviewed this incident and the safety report. it is acknowledged that the stringent checking by the perfusion team ensured that the leak was noted and the risk was appropriately managed. the Eurosets report outlines that they are in adherence to the appropriate controls and have met their risk matrix. However, the leak does question quality and a recognition by the community that disposable issues need to be assessed with adequate check.

DETAILS OF INJURY TO PATIENT

No injury

TYPE OF DEVICE/MANUFACTURER

Eurosets Adult A.L. one ECMO oxygenator (EU5019)

ACTION TAKEN

The device was removed and retained an sent off to the manufacturer for inspection.

RESPONSE FROM MANUFACTURER

Full report completed and actions taken.

Eurosets testing:

We set-up the circuit: the reservoir is filled with 500 ml of physiological water; the oxygenator module is filled form the reservoir using a peristaltic pump (Q = 6 l/min) and remained in loop for 10 minutes. We notice a slow dripping in the lower part of the oxygenator, in front of the arterial blood out. To identify the exact point of leakage, the device was sectioned removing the Gas-escape lid, refilled with water, and then pressurized.

The point of loss occurred due to the fiber breakage of the last winding at the blood outlet.

The leak reported during the priming phase can be related to a fiber rupture occurred after Eurosets manufacturing and test phases. however, further improvement processes have been implemented to reduce this occurrence

ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
059
1/5/20
Failure to remove CO2 effectively
REPORT NO
058
1/7/20
Centrifugal Head Issue 3
REPORT NO
057
1/6/20
Centrifugal Head Issue 2
REPORT NO
056
1/1/20
Centrifugal Head Issue
REPORT NO
055
1/2/20
Faulty Pressure Relief Valves
REPORT NO
054
1/4/20
Leaking ALF
REPORT NO
053
1/2/20
Arterial Line Filter Leak
REPORT NO
052
1/3/19
Gas Line Filter Issue
REPORT NO
051
1/12/19
Dual Lumen Cannula Movement
REPORT NO
050
1/10/19
Venous Cannula Damaged
REPORT NO
049
1/12/19
Oxygenation Issue
REPORT NO
048
1/8/19
Roller Pump Failure
REPORT NO
047
1/12/19
Broken Cannula Dilator
REPORT NO
046
6/12/19
Gas Blender Error
REPORT NO
045
20/11/19
Tubing Detachments
REPORT NO
044
27/11/19
Interruption to IABP counterpulsation
REPORT NO
043
1/7/19
Getinge Mal-Position Of IABP Catheter
REPORT NO
042
1/2/19
Avalon Dual Lumen Cannula Crack
REPORT NO
041
1/2/19
Custodial Rubber Fragmentation In Spiked Bag
REPORT NO
040
1/10/18
Reservoir Clot
REPORT NO
039
1/10/18
Pall Clot In Cardioplegia Circuit
REPORT NO
038
1/10/18
Maquet HL20 Cardioplegia Pump Error
REPORT NO
037
1/10/17
Temed D200 Haemoconcentrator Clot
REPORT NO
036
1/10/17
Oxygenator issue
REPORT NO
035
1/10/17
Medtronic Fusion Oxygenator Blood To Gas Leak
REPORT NO
034
Possible Medtronic Fusion Oxygenator Failure
REPORT NO
033
Fannin Heparin Ampules Shattering
REPORT NO
032
Sucker Placed In Raceway In Wrong Direction
REPORT NO
031
High Transmembrane Pressure Gradients
REPORT NO
030
Faulty Connector On ECMO Soft Shell Reservoir Bag
Perfusion Report cards are shown in batches of 30 at any one time where available.
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.