PERFUSION REPORTS

REPORT
016
DATE
TITLE
Additional Sucker Incorrectly Placed In Raceway
DETAILS OF INCIDENT

Trainee Perfusionist added an extra sucker to a pre built circuit & failed to complete the 3 perfusion sucker checks. The 4th and final sucker check, a visual wet test of the suckers with the scrub nurse was only completed for 2 of the 3 suckers i.e. not the additional sucker.

The additional sucker was used as a vent and after x clamping the surgeon asked for the vent to be turned on and noted it was blowing not sucking, this was switched off and corrected. The surgeon was adamant that the Aorta was X clamped before the vent was switched on and that no air entered the patient.

DETAILS OF INJURY TO PATIENT

Post operatively the patient showed Lt sided stroke and the family was informed of the incident involving the reversed suction tubing.

The patient and family have been kept fully informed as to the possible reasons for the stroke - and that stroke is a known risk from cardiac surgery and cardiopulmonary bypass.

The family has also been kept fully informed of the mistakes made during setup and checking of the suckers.

TYPE OF DEVICE/MANUFACTURER

N/A

ACTION TAKEN

A Hospital Incident Report was filed on the day of the incident. A panel met and discussed the incident and decided to conduct a full SUI investigation with outcome of investigation sent to family and the SHA.

A full review of practices has been undertaken, this has resulted in a number of changes to procedure and/or protocols.

RESPONSE FROM MANUFACTURER

N/A

ADDTIONAL INFORMATION

N/A

SUPPORTING VISUALS/VIDEO

N/A

SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
120
1/5/24
Disposable Incident
REPORT NO
119
Hardware: HCU Contamination
REPORT NO
118
12/9/23
Human Factors: ECMO Mobilisation
REPORT NO
117
27/9/23
Disposables: Oxygenator
REPORT NO
116
14/9/23
Disposable Issue: Reservoir in CPB circuit
REPORT NO
115
13/9/23
Disposable Error
REPORT NO
114
13/7/23
Dispoables
REPORT NO
112
20/2/23
HLM Hardware Error: S5 HLM - Dual roller cardioplegia
REPORT NO
111
10/2/23
Human Factors
REPORT NO
110
31/1/23
DISPOSABLE FAULT
REPORT NO
109
19/1/23
Disposable issue
REPORT NO
108
3/1/23
Disposable issue: Oxygenator leak
REPORT NO
107
6/12/22
Hardware fault. Electronic gas Blender failure
REPORT NO
106
1/11/22
ECG Interference
REPORT NO
105
9/9/22
Issue with use of disposable
REPORT NO
104
13/6/22
Human Factors
REPORT NO
103
25/7/22
Human Factors
REPORT NO
102
7/7/22
Disposable issue
REPORT NO
101
6/4/21
Irregular blood flow/foreign looking substance in Oxygenator
REPORT NO
100
20/4/22
Gas Line Filter Fault
REPORT NO
099
20/4/22
Human Factors: BiVAD Error
REPORT NO
098
4/4/22
REPORT NO
097
19/3/22
Disposables Issue
REPORT NO
096
18/3/22
Bubbles Sensor Activation
REPORT NO
095
1/1/22
Protamine Administration
REPORT NO
094
1/11/21
Oxygenation failure upon commencement of CPB
REPORT NO
093
1/11/21
Detached bullet tip of sump-updated response Sep 2022
REPORT NO
092
1/10/21
Vapourizer Issue
REPORT NO
091
1/10/21
Oyxgenator Leak
REPORT NO
090
1/10/21
Breach of chemical cleaner
Perfusion Report cards are shown in batches of 30 at any one time where available.
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.