PERFUSION REPORTS

REPORT
018
DATE
TITLE
Medos Heat Exchanger Leak
DETAILS OF INCIDENT

During cardiopulmonary bypass (MVR+CABG) blood was seen in the water lines that connect the heater cooler unit to the oxygenator’s heat exchanger. This obviously suggests a leak in the heat exchanger and the patients’ blood and water could be potentially mixing. The water lines were immediately switched off and disconnected, and the consultant surgeon informed.
The recommended procedure is to change out the oxygenator following heat exchanger leak. However, the consultant cardiac surgeon felt that this was not a safe option at the time as the oxygenator was still working, so we remained on bypass. The cross clamp was removed and the patient was rewarmed from 28°C to 34°C using the separate heat exchanger on the cardioplegia circuit and two fluid warmers donated by the anaesthetic team. Once the patient reached 34°C, the patient was ventilated and bypass was terminated and the oxygenator replaced with a new one. Bypass was re-commenced and the patient fully rewarmed. Bypass was discontinued and the patient was stable. It should be noted that a full heat exchanger leak test was performed before the bypass circuit was primed and no fault was detected (water circulated at 37oC for 5mins prior to adding prime solution, no pressure test).

DETAILS OF INJURY TO PATIENT

Unknown. Patient stable and extubated the following morning.

TYPE OF DEVICE/MANUFACTURER

N/A

ACTION TAKEN

Microbiology was informed and antibiotics were given. A water sample from the heater cooler unit was sent to the lab for investigation, and the oxygenator was sent to Medos. A Datix report has been completed and MHRA informed by Medos.

RESPONSE FROM MANUFACTURER

N/A

ADDTIONAL INFORMATION

N/A

SUPPORTING VISUALS/VIDEO

N/A

SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
120
1/5/24
Disposable Incident
REPORT NO
119
Hardware: HCU Contamination
REPORT NO
118
12/9/23
Human Factors: ECMO Mobilisation
REPORT NO
117
27/9/23
Disposables: Oxygenator
REPORT NO
116
14/9/23
Disposable Issue: Reservoir in CPB circuit
REPORT NO
115
13/9/23
Disposable Error
REPORT NO
114
13/7/23
Dispoables
REPORT NO
112
20/2/23
HLM Hardware Error: S5 HLM - Dual roller cardioplegia
REPORT NO
111
10/2/23
Human Factors
REPORT NO
110
31/1/23
DISPOSABLE FAULT
REPORT NO
109
19/1/23
Disposable issue
REPORT NO
108
3/1/23
Disposable issue: Oxygenator leak
REPORT NO
107
6/12/22
Hardware fault. Electronic gas Blender failure
REPORT NO
106
1/11/22
ECG Interference
REPORT NO
105
9/9/22
Issue with use of disposable
REPORT NO
104
13/6/22
Human Factors
REPORT NO
103
25/7/22
Human Factors
REPORT NO
102
7/7/22
Disposable issue
REPORT NO
101
6/4/21
Irregular blood flow/foreign looking substance in Oxygenator
REPORT NO
100
20/4/22
Gas Line Filter Fault
REPORT NO
099
20/4/22
Human Factors: BiVAD Error
REPORT NO
098
4/4/22
REPORT NO
097
19/3/22
Disposables Issue
REPORT NO
096
18/3/22
Bubbles Sensor Activation
REPORT NO
095
1/1/22
Protamine Administration
REPORT NO
094
1/11/21
Oxygenation failure upon commencement of CPB
REPORT NO
093
1/11/21
Detached bullet tip of sump-updated response Sep 2022
REPORT NO
092
1/10/21
Vapourizer Issue
REPORT NO
091
1/10/21
Oyxgenator Leak
REPORT NO
090
1/10/21
Breach of chemical cleaner
Perfusion Report cards are shown in batches of 30 at any one time where available.
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.