PERFUSION REPORTS

REPORT
052
DATE
1/3/2019
TITLE
Gas Line Filter Issue
DETAILS OF INCIDENT

On-going problem with Liva Nova green gas line with filter since November 2018 (at least 10 gas lines that reported department are aware of).  

Noticed pre-CPB when checking gas flow through gas line, pressure readings over 30mmHg on Maquet electronic gas blender.  When gas sweep increased it alarmed at 50mmHg.  When the gas line filter was disconnected the pressure returned to 0.

This has happened numerous times when using different Maquet electronic gas blenders (EGB 40).  It was not noted when using Sechrist blenders as pressure readings are not available on these blenders.  

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

Liva Nova gas line

Product Code: CP8049

Ref: 08049

ACTION TAKEN

Reported to Liva Nova. 6 gas lines that were deemed faulty were collected by Liva Nova for further investigation.

Liva Nova have supplied an alternative gas line filter (Pall) in case further issues arise while using the Maquet electronic gasblender.  When using the Sechrist blenders and the Liva Nova custom pack the gas line is automatically swapped for a Pall gas line.  

RESPONSE FROM MANUFACTURER

See attached below

ADDTIONAL INFORMATION

Upon reflection what do you think went well?

The issues was always identified prior to going on CPB so no patient was put at risk.  Members of the perfusion team were informed quickly so all were aware and able to keep defective items.  It was reported quickly to Liva Nova.  

The team reflected on the use of gas blenders without the ability to notice pressure changes within the line.  

New action plans/changes as a result of the event?

Liva Nova custom packs going forward will only be supplied with Pall gas line filter.  

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
089
1/9/21
Sechrist Gas Blender Sweep on ECMO
REPORT NO
088
1/9/21
Oxygenator High Pressure Events
REPORT NO
087
1/10/21
Leaking one way valve
REPORT NO
086
1/7/21
Electronic Gas Blender Fault
REPORT NO
085
1/2/21
Protamine Administration Issue
REPORT NO
084
1/6/21
Damaged roller pump lid
REPORT NO
083
1/6/21
IABP malfunction
REPORT NO
082
1/1/21
Accidental cardioplegia administration
REPORT NO
081
1/4/21
S5 Double Roller Pump incident
REPORT NO
080
1/5/21
Human factors issue
REPORT NO
079
1/6/21
Centrimag battery module failure
REPORT NO
078
1/6/21
Sampling manifold valve fault
REPORT NO
077
1/5/21
IABP transport issue
REPORT NO
076
1/4/21
Modified Ultrafiltration (MUF) Issue
REPORT NO
075
1/3/21
Electromagnetic interference with HLM
REPORT NO
074
1/2/21
HCU40 Failure
REPORT NO
073
1/2/21
HCU40 valve failures
REPORT NO
072
1/1/21
HCU40 failure
REPORT NO
071
1/1/21
LIVA NOVA STATEMENT ON CENTRIFUGAL HEAD
REPORT NO
070
1/1/21
Loose component on circuit
REPORT NO
069
1/12/20
Centrifugal Head noise
REPORT NO
068
1/11/20
Venous Reservoir disposable issue
REPORT NO
067
1/11/20
Failure of Heater Cooler due to leaking valves
REPORT NO
066
1/11/20
Failure of Valves on heater cooler unit
REPORT NO
065
1/9/20
Intra Aortic Balloon Pump (IABP) Failure
REPORT NO
064
1/9/20
Centrifugal head failure
REPORT NO
063
1/8/20
Centrifugal head issue 4
REPORT NO
062
1/8/20
Human Factors Issue
REPORT NO
061
1/8/20
LIVA NOVA RESPONSE TO REPORTED CENTRIFUGAL HEAD ISSUE
REPORT NO
060
1/8/20
Oxygenator Fibre Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.