PERFUSION REPORTS

REPORT
052
DATE
1/3/2019
TITLE
Gas Line Filter Issue
DETAILS OF INCIDENT

On-going problem with Liva Nova green gas line with filter since November 2018 (at least 10 gas lines that reported department are aware of).  

Noticed pre-CPB when checking gas flow through gas line, pressure readings over 30mmHg on Maquet electronic gas blender.  When gas sweep increased it alarmed at 50mmHg.  When the gas line filter was disconnected the pressure returned to 0.

This has happened numerous times when using different Maquet electronic gas blenders (EGB 40).  It was not noted when using Sechrist blenders as pressure readings are not available on these blenders.  

DETAILS OF INJURY TO PATIENT

None

TYPE OF DEVICE/MANUFACTURER

Liva Nova gas line

Product Code: CP8049

Ref: 08049

ACTION TAKEN

Reported to Liva Nova. 6 gas lines that were deemed faulty were collected by Liva Nova for further investigation.

Liva Nova have supplied an alternative gas line filter (Pall) in case further issues arise while using the Maquet electronic gasblender.  When using the Sechrist blenders and the Liva Nova custom pack the gas line is automatically swapped for a Pall gas line.  

RESPONSE FROM MANUFACTURER

See attached below

ADDTIONAL INFORMATION

Upon reflection what do you think went well?

The issues was always identified prior to going on CPB so no patient was put at risk.  Members of the perfusion team were informed quickly so all were aware and able to keep defective items.  It was reported quickly to Liva Nova.  

The team reflected on the use of gas blenders without the ability to notice pressure changes within the line.  

New action plans/changes as a result of the event?

Liva Nova custom packs going forward will only be supplied with Pall gas line filter.  

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
059
1/5/20
Failure to remove CO2 effectively
REPORT NO
058
1/7/20
Centrifugal Head Issue 3
REPORT NO
057
1/6/20
Centrifugal Head Issue 2
REPORT NO
056
1/1/20
Centrifugal Head Issue
REPORT NO
055
1/2/20
Faulty Pressure Relief Valves
REPORT NO
054
1/4/20
Leaking ALF
REPORT NO
053
1/2/20
Arterial Line Filter Leak
REPORT NO
052
1/3/19
Gas Line Filter Issue
REPORT NO
051
1/12/19
Dual Lumen Cannula Movement
REPORT NO
050
1/10/19
Venous Cannula Damaged
REPORT NO
049
1/12/19
Oxygenation Issue
REPORT NO
048
1/8/19
Roller Pump Failure
REPORT NO
047
1/12/19
Broken Cannula Dilator
REPORT NO
046
6/12/19
Gas Blender Error
REPORT NO
045
20/11/19
Tubing Detachments
REPORT NO
044
27/11/19
Interruption to IABP counterpulsation
REPORT NO
043
1/7/19
Getinge Mal-Position Of IABP Catheter
REPORT NO
042
1/2/19
Avalon Dual Lumen Cannula Crack
REPORT NO
041
1/2/19
Custodial Rubber Fragmentation In Spiked Bag
REPORT NO
040
1/10/18
Reservoir Clot
REPORT NO
039
1/10/18
Pall Clot In Cardioplegia Circuit
REPORT NO
038
1/10/18
Maquet HL20 Cardioplegia Pump Error
REPORT NO
037
1/10/17
Temed D200 Haemoconcentrator Clot
REPORT NO
036
1/10/17
Oxygenator issue
REPORT NO
035
1/10/17
Medtronic Fusion Oxygenator Blood To Gas Leak
REPORT NO
034
Possible Medtronic Fusion Oxygenator Failure
REPORT NO
033
Fannin Heparin Ampules Shattering
REPORT NO
032
Sucker Placed In Raceway In Wrong Direction
REPORT NO
031
High Transmembrane Pressure Gradients
REPORT NO
030
Faulty Connector On ECMO Soft Shell Reservoir Bag
Perfusion Report cards are shown in batches of 30 at any one time where available.
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.