PERFUSION REPORTS

REPORT
108
DATE
3/1/2023
TITLE
Disposable issue: Oxygenator leak
DETAILS OF INCIDENT

During standard priming checks there was a leak noted from the housing of the oxygenator. The heater cooler was emptied to ensure there wasn't a leak from the heater cooler. The oxygenator was pressurised with recirculation lines clamped and the inlet to the oxygenator also clamped (to isolate the pump boot) and the leak persisted with a constant pressure drop noted in the oxygenator.

DETAILS OF INJURY TO PATIENT

No injury.

TYPE OF DEVICE/MANUFACTURER

Liva Nova Oxygenator product code: IN01108H

ACTION TAKEN

A replacement set was utilised and the faulty set was put aside for analysis by the manufacturer, manufacturer was informed and complaint started. Ulysses (datix) completed for trust tracking.

RESPONSE FROM MANUFACTURER

Returned part (SN 592786-0092) was visually inspected and no damaged component was identified.

A subsequent dye-leak test on blood side was carried out to verify the integrity of the device.

The unit was filled with methylene blue and pressurized by sealing all the outlet ports, in line with the maximum internal pressure value permitted in blood path (1 Bar) as per product specifications.

The internal pressure value was kept monitored with a pressure gauge connected to the circuit at the level of the inlet.

In the early filling phase, the inner core of oxygenator fiber bundle was found to be broken (it turned blue) and spillage of dye solution from open slot region of the device was reproduced.

Based on observed findings, reported failure was confirmed and traced back to damaged fiber/s of the oxygenator module.

In such a case, fluid (priming or blood) enters into the ruptured capillary and leaks out of gas escape compartment following gas flow direction.

Considering that 100% of the units in production line is subjected to final leak test and that involved unit successfully passed the in-process check and no trend has been registered for similar cases, it cannot be excluded that fiber bundle capillary manufacturing variability with contribution of sterilization thermal and/or transportation mechanical stress could have led to breakage during use.

No specific action is currently deemed as necessary.

LivaNova maintains and documents periodic customer events monitoring process in order to evaluate actions for products improvement.

ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
059
1/5/20
Failure to remove CO2 effectively
REPORT NO
058
1/7/20
Centrifugal Head Issue 3
REPORT NO
057
1/6/20
Centrifugal Head Issue 2
REPORT NO
056
1/1/20
Centrifugal Head Issue
REPORT NO
055
1/2/20
Faulty Pressure Relief Valves
REPORT NO
054
1/4/20
Leaking ALF
REPORT NO
053
1/2/20
Arterial Line Filter Leak
REPORT NO
052
1/3/19
Gas Line Filter Issue
REPORT NO
051
1/12/19
Dual Lumen Cannula Movement
REPORT NO
050
1/10/19
Venous Cannula Damaged
REPORT NO
049
1/12/19
Oxygenation Issue
REPORT NO
048
1/8/19
Roller Pump Failure
REPORT NO
047
1/12/19
Broken Cannula Dilator
REPORT NO
046
6/12/19
Gas Blender Error
REPORT NO
045
20/11/19
Tubing Detachments
REPORT NO
044
27/11/19
Interruption to IABP counterpulsation
REPORT NO
043
1/7/19
Getinge Mal-Position Of IABP Catheter
REPORT NO
042
1/2/19
Avalon Dual Lumen Cannula Crack
REPORT NO
041
1/2/19
Custodial Rubber Fragmentation In Spiked Bag
REPORT NO
040
1/10/18
Reservoir Clot
REPORT NO
039
1/10/18
Pall Clot In Cardioplegia Circuit
REPORT NO
038
1/10/18
Maquet HL20 Cardioplegia Pump Error
REPORT NO
037
1/10/17
Temed D200 Haemoconcentrator Clot
REPORT NO
036
1/10/17
Oxygenator issue
REPORT NO
035
1/10/17
Medtronic Fusion Oxygenator Blood To Gas Leak
REPORT NO
034
Possible Medtronic Fusion Oxygenator Failure
REPORT NO
033
Fannin Heparin Ampules Shattering
REPORT NO
032
Sucker Placed In Raceway In Wrong Direction
REPORT NO
031
High Transmembrane Pressure Gradients
REPORT NO
030
Faulty Connector On ECMO Soft Shell Reservoir Bag
Perfusion Report cards are shown in batches of 30 at any one time where available.
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.