PERFUSION REPORTS

REPORT
150
DATE
21/10/2026
TITLE
Governance issue of undocumented IABP therapy
DETAILS OF INCIDENT

A 75-year-old man underwent an aortic valve replacement and CABG. The operation was complex and an intra-aortic balloon pump (IABP) in situ. Over the following hours, the balloon pump intermittently alarmed for gas leak. Reviews and troubleshooting from anaesthetic, surgical, and perfusion teams occurred, and the IABP was removed during a subsequent chest washout. Shortly after removal, the patient was found to have signs of acute right leg limb ischaemia. A CT scan confirmed gas embolism in the right iliac and distal vessels. He later required a right lower limb amputation.

DETAILS OF INJURY TO PATIENT

Lower leg amputation to patient.

TYPE OF DEVICE/MANUFACTURER
ACTION TAKEN
RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION

What happened?

‍

  • The IABP was inserted intraoperatively, the following day, the IABP console began intermittently alarming indicating a possible gas leak.

‍

  • This was escalated and was reviewed at the bedside by the perfusionist. Standard checks were conducted by the perfusionist, and no additional evidence of a significant leak/ rupture was noted. A second perfusionist reviewed later that evening.

‍

  • The perfusionist did raise with the clinical team whether the IABP was still required.

‍

  • A second perfusionist also conducted trouble shooting and swapped out the console in case that was the cause. This appeared to resolve the alarm issue at the time.

‍

  • Overnight, this IABP console continued to alarm.

‍

  • During morning review with perfusion, surgical and anaesthetic teams, blood was noted in the line at this stage for the first time, indicating possible rupture of the IABP balloon, and a decision was taken to remove the IABP.

‍

  • Removal of the IAPB was noted to be straightforward. Nil concerns noted at the time of removal. Pressure applied to the area following removal. The balloon tip was discarded and not inspected.

‍

What was expected?

‍

  • Clear documentation of all care and intervention delivered to the patient.

‍

Why was there a difference?

‍

  • Although troubleshooting appears to have been undertaken appropriately (console swap, inspection, review by on‑call perfusion), there is no clear standardised protocol for managing recurrent IABP alarms.

‍

  • Perfusionists do not document actions in the patient’s clinical notes (Cerner electronic record) or any other location, which limited visibility of troubleshooting, communication between the multi-professional team, and possibly reduced continuity of care.
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
091
1/10/21
Oyxgenator Leak
REPORT NO
090
1/10/21
Breach of chemical cleaner
REPORT NO
089
1/9/21
Sechrist Gas Blender Sweep on ECMO
REPORT NO
088
1/9/21
Oxygenator High Pressure Events
REPORT NO
087
1/10/21
Leaking one way valve
REPORT NO
086
1/7/21
Electronic Gas Blender Fault
REPORT NO
085
1/2/21
Protamine Administration Issue
REPORT NO
084
1/6/21
Damaged roller pump lid
REPORT NO
083
1/6/21
IABP malfunction
REPORT NO
082
1/1/21
Accidental cardioplegia administration
REPORT NO
081
1/4/21
S5 Double Roller Pump incident
REPORT NO
080
1/5/21
Human factors issue
REPORT NO
079
1/6/21
Centrimag battery module failure
REPORT NO
078
1/6/21
Sampling manifold valve fault
REPORT NO
077
1/5/21
IABP transport issue
REPORT NO
076
1/4/21
Modified Ultrafiltration (MUF) Issue
REPORT NO
075
1/3/21
Electromagnetic interference with HLM
REPORT NO
074
1/2/21
HCU40 Failure
REPORT NO
073
1/2/21
HCU40 valve failures
REPORT NO
072
1/1/21
HCU40 failure
REPORT NO
071
1/1/21
LIVA NOVA STATEMENT ON CENTRIFUGAL HEAD
REPORT NO
070
1/1/21
Loose component on circuit
REPORT NO
069
1/12/20
Centrifugal Head noise
REPORT NO
068
1/11/20
Venous Reservoir disposable issue
REPORT NO
067
1/11/20
Failure of Heater Cooler due to leaking valves
REPORT NO
066
1/11/20
Failure of Valves on heater cooler unit
REPORT NO
065
1/9/20
Intra Aortic Balloon Pump (IABP) Failure
REPORT NO
064
1/9/20
Centrifugal head failure
REPORT NO
063
1/8/20
Centrifugal head issue 4
REPORT NO
062
1/8/20
Human Factors Issue
Perfusion Report cards are shown in batches of 30 at any one time where available.
151
150
149
148
147
146
145
144
143
142
141
140
139
138
137
136
135
134
133
132
131
130
129
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.