Hardware Error: Bubble Detector
DETAILS OF INCIDENT

While on bypass with a warm patient post XC removal the bubble detector alarmed and the alarm could not be cleared. This occured on a centrifugal pump, so the ERC automatically clamped the arterial line. As the alarm could not be cleared the ERC would not open.

DETAILS OF INJURY TO PATIENT
TYPE OF DEVICE/MANUFACTURER
ACTION TAKEN

The Arterial line was checked for air, none was found. The alarm could not be cleared. The bubble detector was moved to a different point along the tubing and the alarm could still not be cleared. The bubble alarm was turned off and on twice without resolution. Hand cranking took place while a colleague was called to assist. Colleague switched off the alarm, moved the pump from hand crank back into the motor and successfully went back on bypass. The bubble alarm was turned back on after about 5 minutes when the patient was stable. No more issues.

RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
001
1/1/09
Oxygenator Water To Blood Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
151
150
149
148
147
146
145
144
143
142
141
140
139
138
137
136
135
134
133
132
131
130
129
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.