PERFUSION REPORTS

REPORT
126
DATE
1/3/2025
TITLE
Oxygenator Failure
DETAILS OF INCIDENT

Upon commencing bypass, it was noticed that the CDIinline blood gas monitoring device was showing a lower pO2 than expected(around 17), with an FiO2 set at 75%. Gases were taken to calibrate the device,which showed it was accurate, so Fi02 was increased to 90%. ACT was checked and came back at 365, so 10,000iu of heparin was given and a repeat ACT gave aresult of mid 400's.

During this time, the cross-clamp had been applied,cardioplegia given and the first anastomosis bypass graft had commenced.The pre-membrane pressure alarmed, indicating an oxygenator issue, so another perfusionist was called to assist, anaesthetist was informed that there may be an oxygenator issue, whilst checks were made to see if anything obvious couldbe wrong, no line kinks were found, and the gas line was changed out. It was determined that there was a problem with the oxygenator as the pre-membrane pressure continued to increase, post-membrane pressure decreased, and pO2 levels were also decreasing, so the surgeon was informed. It was decided that we would remove the cross-clamp, stay at 32 degrees and come offbypass to then change out the oxygenator. The pre-membrane pressure was preventing full flow (5.3L/min), and the patients MAP was in the low 30's, despite giving a large amount of 100mc/ml Phenylephrine. This pressure alarm was overridden so we could achieve full flow on bypass and gain stability to then come off. NIRS were also then applied to monitor head saturation, which were in the 60's and remained so throughout therest of the operation .Once off bypass, 3 perfusionists worked together to change out the oxygenator,de-air the circuit and commence bypass. This was done in 6 minutes.

Bypass was commenced, full flow achieved, pre and post-membrane pressures within normal range, and pO2 also within normal range with an acceptable FiO2. Blood gases were taken to calibrate the CDI, and all agreed to continue as planned. Cross-clamp reapplied, cardioplegia given and grafting commenced. Upon rewarming, pO2 levels started to decrease and were not greatly improving when increasing FiO2. When the cross-clamp was removed, I had reached using an FiO2of 90%, the anaesthetist was informed so we could prepare to put the lungs onas soon as possible, and the surgeons were then also informed. The lungs were put on to finish the top ends of each anastomosis with a view of coming off bypass as soon as possible. Bypass was ended with FiO2 running at 100% and a pO2 of 13 Both oxygenators have been retained to be sent to the manufacturer for testing. Patient had no comorbidities prior to surgery or any known coagulopathy issues. First bypass run was 54 mins, second bypass run 74 mins.

  • The second oxy did have an increasing transmembrane pressure but not to the same extent as the first.
  • Gas flow was checked and attached directly to test flow which was adequated so no gas issue.
  • Baseline ACT was over 400, the first ACT on bypass was 360, and then ACT’s remained therapeutic for the rest of the case (>400), so FFP  wasn’t considered.
  • There was no baseline TEG done for this patient so unable to ascertain coagulopathy.
  • Patient has now been extubated with no ill effects
DETAILS OF INJURY TO PATIENT

No injury

TYPE OF DEVICE/MANUFACTURER

Medtronic

ACTION TAKEN

Oxygenators have been collected and are being tested. .

RESPONSE FROM MANUFACTURER

Awaiting Response.

ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
068
1/11/20
Venous Reservoir disposable issue
REPORT NO
067
1/11/20
Failure of Heater Cooler due to leaking valves
REPORT NO
066
1/11/20
Failure of Valves on heater cooler unit
REPORT NO
065
1/9/20
Intra Aortic Balloon Pump (IABP) Failure
REPORT NO
064
1/9/20
Centrifugal head failure
REPORT NO
063
1/8/20
Centrifugal head issue 4
REPORT NO
062
1/8/20
Human Factors Issue
REPORT NO
061
1/8/20
LIVA NOVA RESPONSE TO REPORTED CENTRIFUGAL HEAD ISSUE
REPORT NO
060
1/8/20
Oxygenator Fibre Leak
REPORT NO
059
1/5/20
Failure to remove CO2 effectively
REPORT NO
058
1/7/20
Centrifugal Head Issue 3
REPORT NO
057
1/6/20
Centrifugal Head Issue 2
REPORT NO
056
1/1/20
Centrifugal Head Issue
REPORT NO
055
1/2/20
Faulty Pressure Relief Valves
REPORT NO
054
1/4/20
Leaking ALF
REPORT NO
053
1/2/20
Arterial Line Filter Leak
REPORT NO
052
1/3/19
Gas Line Filter Issue
REPORT NO
051
1/12/19
Dual Lumen Cannula Movement
REPORT NO
050
1/10/19
Venous Cannula Damaged
REPORT NO
049
1/12/19
Oxygenation Issue
REPORT NO
048
1/8/19
Roller Pump Failure
REPORT NO
047
1/12/19
Broken Cannula Dilator
REPORT NO
046
6/12/19
Gas Blender Error
REPORT NO
045
20/11/19
Tubing Detachments
REPORT NO
044
27/11/19
Interruption to IABP counterpulsation
REPORT NO
043
1/7/19
Getinge Mal-Position Of IABP Catheter
REPORT NO
042
1/2/19
Avalon Dual Lumen Cannula Crack
REPORT NO
041
1/2/19
Custodial Rubber Fragmentation In Spiked Bag
REPORT NO
040
1/10/18
Reservoir Clot
REPORT NO
039
1/10/18
Pall Clot In Cardioplegia Circuit
Perfusion Report cards are shown in batches of 30 at any one time where available.
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.