REPORT
128
DATE
25/5/2025
TITLE
Single Station Selectatec Type Backbar for Vapouriser. 9000-001
DETAILS OF INCIDENT

It involved a SingleStation Selectatec Type Backbar, a mounting unit for Sevoforane Vapourisers. It was custom built for us by a UK Company, supplied via an Irish representative company.

No vapour was being delivered to the patient while on Bypass. The Anaesthetic registrar noticed the BIS number climbing and questioned the level of anaesthesia. He reverted to IV sedation when he became suspicious. Ironically, it was team changeover time of year for anaesthtic registratrs and an inexperienced registrar was thought to be the cause. However, when reported the Hospital engineer thought there was an internal problem. The engineer rigged the gas sampling line of the Anaesthetic monitor to the Oxygenator. That normally displays the percentage of vapour that the patient is receiving from the Anaesthetic via the ventilator, and noted no vapour. Further review noted that the vaporiser valve was fitted incorrecty and worked in reverse. So, as you increased Percentage vapour, the valve closes accordingly, not opens. This was very evident with theGas sampling line.  

These were 4 new Backbars ,built for us by the company outsourced the build to a UK company. Hence their email, attached to the Safety report.

The concern is that otherunits may have had these built from the same company and have not yet discovered the problem.

 

DETAILS OF INJURY TO PATIENT
TYPE OF DEVICE/MANUFACTURER
ACTION TAKEN

Reported to the Society and the Health Products Regulatory Authority (HPRA) Ireland. In accordance to the regulatory obligations under MDR (EU 2017/745) which requires medical devices distributors to conduct post market surveillance and report safety and performance issues. Oxygen care informed

RESPONSE FROM MANUFACTURER

Awaiting response.

ADDTIONAL INFORMATION

Connect the gas sampling line of the Anaesthetic monitor to the gas line of the HLM beforethe patient enters the Theatre to show correct Vapour delivery. This is now part of a regular check off.

 

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
128
25/5/25
Single Station Selectatec Type Backbar for Vapouriser. 9000-001
REPORT NO
127
4/1/25
Eurosets Oxygenator leak
REPORT NO
126
1/3/25
Oxygenator Failure
REPORT NO
125
11/2/25
Hardware Error: Bubble Detector
REPORT NO
124
1/10/24
HCU/ Hardware incident/ Human Factors
REPORT NO
123
5/7/24
LivaNova S5 Roller Pump Fault
REPORT NO
122
5/4/24
Oxygenator Failure
REPORT NO
121
8/3/24
Disposables: Eurosets Adult ECMO Oxygenator
REPORT NO
120
1/5/24
Oxygenator Failure
REPORT NO
119
Hardware: HCU Contamination
REPORT NO
118
12/9/23
Human Factors: ECMO Mobilisation
REPORT NO
117
27/9/23
Disposables: Oxygenator failure
REPORT NO
116
14/9/23
Disposable Issue: Reservoir in CPB circuit
REPORT NO
115
13/9/23
Disposable Error/ Eurosets New Born ECMO oxygenator (EU5056) leak
REPORT NO
114
13/7/23
Dispoables: Bonding Solution on Tubing
REPORT NO
113
21/3/23
ERC Failure
REPORT NO
112
20/2/23
HLM Hardware Error: S5 HLM - Dual roller cardioplegia
REPORT NO
111
10/2/23
Human Factors/ Gas leak
REPORT NO
110
31/1/23
DISPOSABLE FAULT
REPORT NO
109
19/1/23
Disposable issue
REPORT NO
108
3/1/23
Disposable issue: Oxygenator leak
REPORT NO
107
6/12/22
Hardware fault. Electronic gas Blender failure
REPORT NO
106
1/11/22
ECG Interference
REPORT NO
105
9/9/22
Issue with use of disposable
REPORT NO
104
13/6/22
Human Factors
REPORT NO
103
25/7/22
Human Factors
REPORT NO
102
7/7/22
Disposable issue
REPORT NO
101
6/4/21
Irregular blood flow/foreign looking substance in Oxygenator
REPORT NO
100
20/4/22
Gas Line Filter Fault
REPORT NO
099
20/4/22
Human Factors: BiVAD Error
Perfusion Report cards are shown in batches of 30 at any one time where available.
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.