PERFUSION REPORTS

REPORT
129
DATE
3/5/2025
TITLE
Human Factors
DETAILS OF INCIDENT

A patient was transferred on central VA ECMO from critical care to theatre. Piped gas was replaced by an oxygen cylinder for transfer. Upon arriving in theatre the ECMO console alarmed low flow and distracted the perfusionist during plugging the ECMO into mains and swapping the oxygen cylinder for piped gas. It was noted some time later that the patient's pO2 was low on an ABG, and the ECMO lines were dark. It was then noted that the ECMO had been left connected to the oxygen cylinder which had run out.

Immediate action: ECMO connected to piped gas. Pupils checked by anaesthetist and noted to be ok. Patient woke up fine with no harm.

Distraction due to an alarm while transferring gas supply led to human error of one step being missed in the process of connecting to piped gas. Patient woke up and is OK, however, further mitigations are required to reduce the impact of human error.

DETAILS OF INJURY TO PATIENT
TYPE OF DEVICE/MANUFACTURER
ACTION TAKEN

This incident has been added as a risk to the risk register with the following 3 actions:

- Improve the transfer checklist.

- Investigate the option of using oxygen cylinders with low oxygen alarms.

- New ECMO hardware/circuits are planned to be purchased int he future. There are systems on the market that have built in saturation monitoring. It would be ideal if saturation monitoring was a minimum requirement for any new ECMO systems.

RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION

An excellent review and response to a human factor failure that could impact patient care.

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
072
1/1/21
HCU40 failure
REPORT NO
071
1/1/21
LIVA NOVA STATEMENT ON CENTRIFUGAL HEAD
REPORT NO
070
1/1/21
Loose component on circuit
REPORT NO
069
1/12/20
Centrifugal Head noise
REPORT NO
068
1/11/20
Venous Reservoir disposable issue
REPORT NO
067
1/11/20
Failure of Heater Cooler due to leaking valves
REPORT NO
066
1/11/20
Failure of Valves on heater cooler unit
REPORT NO
065
1/9/20
Intra Aortic Balloon Pump (IABP) Failure
REPORT NO
064
1/9/20
Centrifugal head failure
REPORT NO
063
1/8/20
Centrifugal head issue 4
REPORT NO
062
1/8/20
Human Factors Issue
REPORT NO
061
1/8/20
LIVA NOVA RESPONSE TO REPORTED CENTRIFUGAL HEAD ISSUE
REPORT NO
060
1/8/20
Oxygenator Fibre Leak
REPORT NO
059
1/5/20
Failure to remove CO2 effectively
REPORT NO
058
1/7/20
Centrifugal Head Issue 3
REPORT NO
057
1/6/20
Centrifugal Head Issue 2
REPORT NO
056
1/1/20
Centrifugal Head Issue
REPORT NO
055
1/2/20
Faulty Pressure Relief Valves
REPORT NO
054
1/4/20
Leaking ALF
REPORT NO
053
1/2/20
Arterial Line Filter Leak
REPORT NO
052
1/3/19
Gas Line Filter Issue
REPORT NO
051
1/12/19
Dual Lumen Cannula Movement
REPORT NO
050
1/10/19
Venous Cannula Damaged
REPORT NO
049
1/12/19
Oxygenation Issue
REPORT NO
048
1/8/19
Roller Pump Failure
REPORT NO
047
1/12/19
Broken Cannula Dilator
REPORT NO
046
6/12/19
Gas Blender Error
REPORT NO
045
20/11/19
Tubing Detachments
REPORT NO
044
27/11/19
Interruption to IABP counterpulsation
REPORT NO
043
1/7/19
Getinge Mal-Position Of IABP Catheter
Perfusion Report cards are shown in batches of 30 at any one time where available.
133
132
130
129
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.