PERFUSION REPORTS

REPORT
135
DATE
3/2/2026
TITLE
Human Factors
DETAILS OF INCIDENT

During an elective AVR for severe AS, a locum perfusionist was supervising our trainee perfusionist. Bypass was initiated, root vent cannula was placed, and the cross clamp was applied. Cardioplegia delivery was initiated antegradely with cold oxygenated blood cardioplegia. The heart was not arresting, therefore, the aorta was opened and cold oxygenated blood cardioplegia was delivered down the ostia.

DETAILS OF INJURY TO PATIENT

No harm to patient.

TYPE OF DEVICE/MANUFACTURER
ACTION TAKEN

Help was called (2nd perfusionist) to investigate as the locum perfusionist was concerned that the cardioplegia was not effective. After 6 minutes and a total of 1L cold oxygenated cardioplegia the 2nd perfusionist checked the circuit and noticed that the circuit was patent and air free, however, the white clip for the cardioplegia infusion was closed off, therefore the previous delivery had been of cold oxygenated blood only. The clip was subsequently opened and 1L of cold blood cardioplegia was delivered down the ostia and the heart was successfully arrested.

The surgeon and anaesthetist were immediately informed during the fault finding and the issue was resolved. An ABG was done after cardioplegia delivery which confirmed expected normal ABG values for CPB. The rest of the case was uneventful, when the clamp was removed the heart restarted without issue and the ABGs were all as expected.

RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION

‍Staffing did not allow for the senior perfusionist on duty to supervise the trainee as they would normally, as 2 other members of staff were off. The senior perfusionist was on call and would be doing the second case. The staff perfusionist cannot supervise trainees due to departmental educational rules. Therefore, the locum perfusionist (+20 years of experience) was deemed appropriate to supervise the trainee. As a team, there could have perhaps been a second check of the pump since the locum perfusionist is new to the department.

OUTCOME ACTIONS:

  • Review of pre CPB checklist to perhaps include 'cardioplegia clip opened' or completely remove the white clip from the circuit design.
  • Perhaps new practice rule i.e. new locum perfusionists not to supervise trainees unless they have been here more than once.
  • 2 person pre CPB checklist for locum staff or new staff

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
107
6/12/22
Hardware fault. Electronic gas Blender failure
REPORT NO
106
1/11/22
ECG Interference
REPORT NO
105
9/9/22
Issue with use of disposable
REPORT NO
104
13/6/22
Human Factors
REPORT NO
103
25/7/22
Human Factors
REPORT NO
102
7/7/22
Disposable issue
REPORT NO
101
6/4/21
Irregular blood flow/foreign looking substance in Oxygenator
REPORT NO
100
20/4/22
Gas Line Filter Fault
REPORT NO
099
20/4/22
Human Factors: BiVAD Error
REPORT NO
098
4/4/22
REPORT NO
097
19/3/22
Disposables Issue
REPORT NO
096
18/3/22
Bubbles Sensor Activation
REPORT NO
095
1/1/22
Protamine Administration
REPORT NO
094
1/11/21
Oxygenation failure upon commencement of CPB
REPORT NO
093
1/11/21
Detached bullet tip of sump-updated response Sep 2022
REPORT NO
092
1/10/21
Vapourizer Issue
REPORT NO
091
1/10/21
Oyxgenator Leak
REPORT NO
090
1/10/21
Breach of chemical cleaner
REPORT NO
089
1/9/21
Sechrist Gas Blender Sweep on ECMO
REPORT NO
088
1/9/21
Oxygenator High Pressure Events
REPORT NO
087
1/10/21
Leaking one way valve
REPORT NO
086
1/7/21
Electronic Gas Blender Fault
REPORT NO
085
1/2/21
Protamine Administration Issue
REPORT NO
084
1/6/21
Damaged roller pump lid
REPORT NO
083
1/6/21
IABP malfunction
REPORT NO
082
1/1/21
Accidental cardioplegia administration
REPORT NO
081
1/4/21
S5 Double Roller Pump incident
REPORT NO
080
1/5/21
Human factors issue
REPORT NO
079
1/6/21
Centrimag battery module failure
REPORT NO
078
1/6/21
Sampling manifold valve fault
Perfusion Report cards are shown in batches of 30 at any one time where available.
137
136
135
134
133
132
131
130
129
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.