PERFUSION REPORTS

REPORT
139
DATE
20/5/2026
TITLE
Faulty pressure relief valve
DETAILS OF INCIDENT

During set up, the suckers on the heart lung machine were checked and double checked as per protocol, including direction and occlusion. All suckers were working as expected at this stage. Each of the suckers (blue, red, green) have a one way pressure relief valve integrated in to the sucker line. As a final check prior to use, the perfusionist and scrub nurse performed a wet sucker check to ensure the sucker is both in the correct direction and is working as expected. The scrub nurse placed the sterile end of the sucker line in to a pot of saline and the perfusionist turned the sucker on, and they both observe the sucker sucking saline back towards the heart lung machine in the correct direction, and not blowing bubbles in to the saline.

The blue sucker did not suck any saline, nor did it blow any bubbles. It just wasn't working.

DETAILS OF INJURY TO PATIENT

No harm.

TYPE OF DEVICE/MANUFACTURER

One-way valve in a LivaNova custom perfusion tubing system pack.

ACTION TAKEN

The first immediate step was to triple check the direction and  occlusion of the blue sucker, and everything was set up correctly, confirming  that there was no set up error.

 

The second step was to inform the surgeon of the issue with the blue sucker, and to use the spare green sucker instead, which was working as  expected.

 

The final step was to cut out the one way pressure relief valve in the  blue sucker to check if this was the problem (blocked or not opening  correctly). The valve was replaced by a standard 1/4 x 1/4 inch connector.  The sucker was double checked again, and a wet test performed. This time the  sucker was working as expected, suggesting that there was a problem with the valve in the blue sucker line.

RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION

Reported to manufacturer and MHRA (Ref 2026/005/013/501/007)

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
080
1/5/21
Human factors issue
REPORT NO
079
1/6/21
Centrimag battery module failure
REPORT NO
078
1/6/21
Sampling manifold valve fault
REPORT NO
077
1/5/21
IABP transport issue
REPORT NO
076
1/4/21
Modified Ultrafiltration (MUF) Issue
REPORT NO
075
1/3/21
Electromagnetic interference with HLM
REPORT NO
074
1/2/21
HCU40 Failure
REPORT NO
073
1/2/21
HCU40 valve failures
REPORT NO
072
1/1/21
HCU40 failure
REPORT NO
071
1/1/21
LIVA NOVA STATEMENT ON CENTRIFUGAL HEAD
REPORT NO
070
1/1/21
Loose component on circuit
REPORT NO
069
1/12/20
Centrifugal Head noise
REPORT NO
068
1/11/20
Venous Reservoir disposable issue
REPORT NO
067
1/11/20
Failure of Heater Cooler due to leaking valves
REPORT NO
066
1/11/20
Failure of Valves on heater cooler unit
REPORT NO
065
1/9/20
Intra Aortic Balloon Pump (IABP) Failure
REPORT NO
064
1/9/20
Centrifugal head failure
REPORT NO
063
1/8/20
Centrifugal head issue 4
REPORT NO
062
1/8/20
Human Factors Issue
REPORT NO
061
1/8/20
LIVA NOVA RESPONSE TO REPORTED CENTRIFUGAL HEAD ISSUE
REPORT NO
060
1/8/20
Oxygenator Fibre Leak
REPORT NO
059
1/5/20
Failure to remove CO2 effectively
REPORT NO
058
1/7/20
Centrifugal Head Issue 3
REPORT NO
057
1/6/20
Centrifugal Head Issue 2
REPORT NO
056
1/1/20
Centrifugal Head Issue
REPORT NO
055
1/2/20
Faulty Pressure Relief Valves
REPORT NO
054
1/4/20
Leaking ALF
REPORT NO
053
1/2/20
Arterial Line Filter Leak
REPORT NO
052
1/3/19
Gas Line Filter Issue
REPORT NO
051
1/12/19
Dual Lumen Cannula Movement
Perfusion Report cards are shown in batches of 30 at any one time where available.
140
139
138
137
136
135
134
133
132
131
130
129
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.