Possible Medtronic Fusion Oxygenator Failure
DETAILS OF INCIDENT

60 year old 80 kg male patient. Blood group B+ve. Previous PCI During coronary artery bypass grafting surgery, once Cardiopulmonary bypass (CPB) commenced, there were very high pre-oxygenator P1 line pressures in the bypass circuit (450mmHg). The oxygenator was a Medtronic Fusion. P2 post-oxygenator pressures were within 'normal' range. All potential causes were eliminated but it persised.

The high P1 pressures were not resolving and the p02 oxygen measurement started to fall gradually, although not compromising patient oxygenation. Clotting levels were within range and no clots were seen in the oxygenator. The surgeon was kept informed and the situation monitored before progressing with operation. The decision was made to change out the oxygenator. The anaesthetist was informed and the patient was ventilated conventionally.

CPB was stopped and the oxygenator changed. CPB recommenced and the same high P1 pressures seen again but with no compromise to oxygenation so bypass was continued and monitored very closely. After 30 minutes, the pressures fell to 'normal' levels and remained that way for remainder of the bypass. It was also noted that patient haematocrit on CPB was high (38%) and following discussion with the anaesthetist the patient was diluted with the addition of crystalloid.

Post CPB and protamine, a distal vein graft needed repair and heparin was administered and CPB recommenced. P1 and P2 were both within 'normal levels'. As high P1 pressures seen with both devices on initial use, this was suggestive of a patient specific incident. No untoward effects were noted in the post-op period.

DETAILS OF INJURY TO PATIENT

N/A

TYPE OF DEVICE/MANUFACTURER

N/A

ACTION TAKEN

First oxygenator returned to Medtronic for evaluation. No fault was found with the oxygenator.

In-house discussion:

Following the incident there was a feeling that perhaps the oxygenator was changed out too soon and high pre oxygenator pressures would have stabilised.


RESPONSE FROM MANUFACTURER

N/A

ADDTIONAL INFORMATION

N/A

SUPPORTING VISUALS/VIDEO

N/A

SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
029
Sorin Inspire 8F Blood Leak From Temperature Port
REPORT NO
028
Sorin Inspire 8F Oxygenator Detached From Holding Arm
REPORT NO
027
Paragon Oxygenator Gas Inlet Leak
REPORT NO
026
Maquet IABP Console Failure
REPORT NO
025
Sorin EOS Gas To Blood Leak Whilst Priming
REPORT NO
024
Chalice Paragon Midi Heat Exchanger Leak Whilst Priming
REPORT NO
023
Chalice Paragon Heat Exchanger Water To Blood Leak
REPORT NO
022
Vent Pump Boot Tubing Inserted In Wrong Way Round
REPORT NO
021
Paragon Oxygenator Heat Exchanger Leak
REPORT NO
020
S3 E62 Pump Error
REPORT NO
019
Medtronic Potential Oxygenator Leak
REPORT NO
018
Medos Heat Exchanger Leak
REPORT NO
017
Avant D903 Has A Dual Chamber Reservoir Leak
REPORT NO
016
Additional Sucker Incorrectly Placed In Raceway
REPORT NO
015
Large Perished Tear Observed On Tubing At Back Of Vaporiser
REPORT NO
014
Gas Blender Flow Control Immobilised
REPORT NO
013
Isoflurane Vaporiser Seal Leak Due To Malposition
REPORT NO
012
1/2/13
Hypercoagulable Disorder May Have Led To Oxygenator Failure
REPORT NO
011
1/1/13
LV Vent Placed In Raceway In Wrong Direction
REPORT NO
010
1/3/12
Sucker Placed In Raceway In Wrong Direction
REPORT NO
009
1/2/12
Similarity In Heparin And St Thomas Cardioplegia Ampules
REPORT NO
008
1/7/11
Aorta Remained Clamped Post CPB
REPORT NO
007
1/5/11
Deficiencies In The QC At Terumo Ann Arbor, USA.
REPORT NO
006
1/4/11
Awareness Of Suckers When Using Bio-Glues
REPORT NO
005
1/2/11
Patient Exsanguinated From Additional Venous Line Post CPB
REPORT NO
004
1/1/11
Possible Inaccuracies In Colloid Research
REPORT NO
003
1/2/10
Difficulties In Visualising Blender Reading
REPORT NO
002
1/1/10
Luer Cap Found In Venous Inlet Port
REPORT NO
001
1/1/09
Oxygenator Water To Blood Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.