PERFUSION REPORTS

REPORT
038
DATE
1/10/2018
TITLE
Maquet HL20 Cardioplegia Pump Error
DETAILS OF INCIDENT

The cardioplegia pump on the heart-lung machine accelerated uncontrollably during cardioplegia administration. Cardioplegia was administered through a twin-pump module on a Maquet HL20, one blood pump (master) and the other cardioplegia solution pump (slave). During this incident cardioplegia was being given via a vein graft to a coronary artery. The risk of an accelerating uncontrolled pump when delivering cardioplegia down this route is to damage the vessel it is being given down and/or damage to the vein graft to coronary artery suture site.

This is the units third such event in twelve months.

The cardioplegia pumps have the following alarms in place at all times:

  • Arterial flow zero: so no cardioplegia flow is possible when there is no arterial flow detected on the HL20.
  • Runaway: where the cardioplegia pumps speed is greater than the revs set. If triggered this should stop the affected pump.
  • Pressure alarm: where a high and low pressure reached via the cardioplegia line pressure (280mmHg and minus10mmHg). If triggered this should stop the master cardioplegia pump (and therefore the cardioplegia slave as well).

In all of the units episodes (three episodes) there has been audible alarms either for Runaway or Pressure or both but the stop feature did not happen and in each episode the main power switch for that twin-pump module had to be turned off for the Runaway pump to be stopped. The same HLM was in use for each event operated by different perfusion scientists.

DETAILS OF INJURY TO PATIENT

None, graft site re-checked by Consultant (SA) and confirmed as intact and undamaged.

TYPE OF DEVICE/MANUFACTURER

N/A

ACTION TAKEN

Firstly, the immediate and extremely swift action taken by the perfusionist prevented significant harm being caused to the patient. the presence of a colleague enabled the surgery to continue without delay. The surgeon was very supportive. There had been no deviation from Clinical Perfusion policy and all alarms had been set correctly. On seeing the analogue pressure gauge turning 'around the clock', thereby indicating a significant pressure increase in the cardioplegia circuit, she clamped the patient delivery line and switched off the main power switch to that twin-pump module. A second perfusionist was immediately called into theatre the decision to change out the twin-pump module from the heart-lung machine was taken. There was no interruption to surgery but a delay of several seconds for a fresh administration of cardioplegia. The surgeon was advised of the delay. Once completed, the surgeon was advised of the detail of the incident and he checked the vein graft to coronary artery suture site for any damage. This he was happy to do and confirmed that the site was undamaged and looked fine. The twin-pump has been quarantined and the manufacturer was contacted.



RESPONSE FROM MANUFACTURER

The pump was thoroughly checked by the servicing engineer, and no fault was identified.

ADDTIONAL INFORMATION

N/A

SUPPORTING VISUALS/VIDEO

N/A

SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
029
Sorin Inspire 8F Blood Leak From Temperature Port
REPORT NO
028
Sorin Inspire 8F Oxygenator Detached From Holding Arm
REPORT NO
027
Paragon Oxygenator Gas Inlet Leak
REPORT NO
026
Maquet IABP Console Failure
REPORT NO
025
Sorin EOS Gas To Blood Leak Whilst Priming
REPORT NO
024
Chalice Paragon Midi Heat Exchanger Leak Whilst Priming
REPORT NO
023
Chalice Paragon Heat Exchanger Water To Blood Leak
REPORT NO
022
Vent Pump Boot Tubing Inserted In Wrong Way Round
REPORT NO
021
Paragon Oxygenator Heat Exchanger Leak
REPORT NO
020
S3 E62 Pump Error
REPORT NO
019
Medtronic Potential Oxygenator Leak
REPORT NO
018
Medos Heat Exchanger Leak
REPORT NO
017
Avant D903 Has A Dual Chamber Reservoir Leak
REPORT NO
016
Additional Sucker Incorrectly Placed In Raceway
REPORT NO
015
Large Perished Tear Observed On Tubing At Back Of Vaporiser
REPORT NO
014
Gas Blender Flow Control Immobilised
REPORT NO
013
Isoflurane Vaporiser Seal Leak Due To Malposition
REPORT NO
012
1/2/13
Hypercoagulable Disorder May Have Led To Oxygenator Failure
REPORT NO
011
1/1/13
LV Vent Placed In Raceway In Wrong Direction
REPORT NO
010
1/3/12
Sucker Placed In Raceway In Wrong Direction
REPORT NO
009
1/2/12
Similarity In Heparin And St Thomas Cardioplegia Ampules
REPORT NO
008
1/7/11
Aorta Remained Clamped Post CPB
REPORT NO
007
1/5/11
Deficiencies In The QC At Terumo Ann Arbor, USA.
REPORT NO
006
1/4/11
Awareness Of Suckers When Using Bio-Glues
REPORT NO
005
1/2/11
Patient Exsanguinated From Additional Venous Line Post CPB
REPORT NO
004
1/1/11
Possible Inaccuracies In Colloid Research
REPORT NO
003
1/2/10
Difficulties In Visualising Blender Reading
REPORT NO
002
1/1/10
Luer Cap Found In Venous Inlet Port
REPORT NO
001
1/1/09
Oxygenator Water To Blood Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
120
119
118
117
116
115
114
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
028
027
026
025
024
023
022
021
020
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.