PERFUSION REPORTS

REPORT
136
DATE
28/1/2026
TITLE
Hardware: ACT machine
DETAILS OF INCIDENT

During an elective procedure that turned into a full day case due to complications and significant bleeding, the patient required a second emergency initiation of CPB. Systemic heparinisation was administered by the anaesthetist. The perfusion team, comprising of the on call perfusionist and an additional perfusionist who remained beyond their scheduled hours added 15,000iu heparin to the CPB circuit. Despite the complex surgical circumstances, the second bypass run was well managed. There were no circuit related complications i.e. elevated arterial line pressures, and the CPB proceeded smoothly. Notably, the ACT remained consistently elevated (>1000 seconds) which is typical in cases of extensive heparinisation.

Upon termination of bypass, the ACT device displayed a 'Heater Error'. The perfusionist suspected that this was either a device malfunction or overheating due to prolonged continuous use. This was also checked by the second perfusionist. Post bypass, the clinical environment remained highly active and busy, with IABP support, ongoing emergency cell salvage, and volume transfusion. Therefore, the perfusionists did not find it appropriate to inform the surgeons and anaesthetist of this error. Protamine was administered per standard protocol at the surgeon's request to reverse heparin effects. Normally, ACT values return close to baseline following reversal, however, in this instance the ACT continued to rise abnormally (>700 seconds). A repeat ACT sample tested on an alternative machine returned a value of 112 seconds, confirming that the intial device was likely faulty or had overheated. As far as I am aware no further protamine was administered and the case ended as expected.

DETAILS OF INJURY TO PATIENT

No harm to patient.

TYPE OF DEVICE/MANUFACTURER

Werfen ACT machine (GEM hemochron 100)

ACTION TAKEN

The faulty ACT machine was sent to the manufacturer for investigation. Machine was repaired and is now back in service.

RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
107
6/12/22
Hardware fault. Electronic gas Blender failure
REPORT NO
106
1/11/22
ECG Interference
REPORT NO
105
9/9/22
Issue with use of disposable
REPORT NO
104
13/6/22
Human Factors
REPORT NO
103
25/7/22
Human Factors
REPORT NO
102
7/7/22
Disposable issue
REPORT NO
101
6/4/21
Irregular blood flow/foreign looking substance in Oxygenator
REPORT NO
100
20/4/22
Gas Line Filter Fault
REPORT NO
099
20/4/22
Human Factors: BiVAD Error
REPORT NO
098
4/4/22
REPORT NO
097
19/3/22
Disposables Issue
REPORT NO
096
18/3/22
Bubbles Sensor Activation
REPORT NO
095
1/1/22
Protamine Administration
REPORT NO
094
1/11/21
Oxygenation failure upon commencement of CPB
REPORT NO
093
1/11/21
Detached bullet tip of sump-updated response Sep 2022
REPORT NO
092
1/10/21
Vapourizer Issue
REPORT NO
091
1/10/21
Oyxgenator Leak
REPORT NO
090
1/10/21
Breach of chemical cleaner
REPORT NO
089
1/9/21
Sechrist Gas Blender Sweep on ECMO
REPORT NO
088
1/9/21
Oxygenator High Pressure Events
REPORT NO
087
1/10/21
Leaking one way valve
REPORT NO
086
1/7/21
Electronic Gas Blender Fault
REPORT NO
085
1/2/21
Protamine Administration Issue
REPORT NO
084
1/6/21
Damaged roller pump lid
REPORT NO
083
1/6/21
IABP malfunction
REPORT NO
082
1/1/21
Accidental cardioplegia administration
REPORT NO
081
1/4/21
S5 Double Roller Pump incident
REPORT NO
080
1/5/21
Human factors issue
REPORT NO
079
1/6/21
Centrimag battery module failure
REPORT NO
078
1/6/21
Sampling manifold valve fault
Perfusion Report cards are shown in batches of 30 at any one time where available.
137
136
135
134
133
132
131
130
129
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.