PERFUSION REPORTS

REPORT
140
DATE
12/6/2026
TITLE
Isoflurane vaporiser
DETAILS OF INCIDENT

Incident date 20/12/2023. Midway through procedure, Isoflurane turned on to treat hypertension. Isoflurane bracket was resting on a sidebar of the HLM. This upward lift broke the seal in the isoflurane, completely cutting gas supply to the oxygenator.

DETAILS OF INJURY TO PATIENT

Patient died post up due to complications.

TYPE OF DEVICE/MANUFACTURER

Datex Ohmeda

ACTION TAKEN

In-line monitoring picked up the problem and raised multiple alarms. 2nd perfusionist came to help. As no clear problem with gas flow due to ball in Sechrist blender floating normally and gas line appeared intact, the decision was made by perfusionist 2, to change oxygenator. Perfusionist 3 came to help with change-out. On recommencing bypass, hypoxia remained, at which point perfusionist 4 removed the vaporiser from the bracket and the issue resolved.

RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION

Full Patient Level Safety Report was carried out. On investigation,  nothing was found to be wrong with the vaporiser or the bracket.  

This is a design fault with the bracket made by G.A.S (General  Anaesthetic Services). I understand this does not have CE marking as I have tried to buy replacements, but cannot do so. MHRA not contacted as there was  no fault found.

 

 

 

Presented this case at the safety committee session at the annual  congress in Nottingham 2025.

 

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
080
1/5/21
Human factors issue
REPORT NO
079
1/6/21
Centrimag battery module failure
REPORT NO
078
1/6/21
Sampling manifold valve fault
REPORT NO
077
1/5/21
IABP transport issue
REPORT NO
076
1/4/21
Modified Ultrafiltration (MUF) Issue
REPORT NO
075
1/3/21
Electromagnetic interference with HLM
REPORT NO
074
1/2/21
HCU40 Failure
REPORT NO
073
1/2/21
HCU40 valve failures
REPORT NO
072
1/1/21
HCU40 failure
REPORT NO
071
1/1/21
LIVA NOVA STATEMENT ON CENTRIFUGAL HEAD
REPORT NO
070
1/1/21
Loose component on circuit
REPORT NO
069
1/12/20
Centrifugal Head noise
REPORT NO
068
1/11/20
Venous Reservoir disposable issue
REPORT NO
067
1/11/20
Failure of Heater Cooler due to leaking valves
REPORT NO
066
1/11/20
Failure of Valves on heater cooler unit
REPORT NO
065
1/9/20
Intra Aortic Balloon Pump (IABP) Failure
REPORT NO
064
1/9/20
Centrifugal head failure
REPORT NO
063
1/8/20
Centrifugal head issue 4
REPORT NO
062
1/8/20
Human Factors Issue
REPORT NO
061
1/8/20
LIVA NOVA RESPONSE TO REPORTED CENTRIFUGAL HEAD ISSUE
REPORT NO
060
1/8/20
Oxygenator Fibre Leak
REPORT NO
059
1/5/20
Failure to remove CO2 effectively
REPORT NO
058
1/7/20
Centrifugal Head Issue 3
REPORT NO
057
1/6/20
Centrifugal Head Issue 2
REPORT NO
056
1/1/20
Centrifugal Head Issue
REPORT NO
055
1/2/20
Faulty Pressure Relief Valves
REPORT NO
054
1/4/20
Leaking ALF
REPORT NO
053
1/2/20
Arterial Line Filter Leak
REPORT NO
052
1/3/19
Gas Line Filter Issue
REPORT NO
051
1/12/19
Dual Lumen Cannula Movement
Perfusion Report cards are shown in batches of 30 at any one time where available.
140
139
138
137
136
135
134
133
132
131
130
129
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.