PERFUSION REPORTS

REPORT
140
DATE
12/6/2026
TITLE
Isoflurane vaporiser
DETAILS OF INCIDENT

Incident date 20/12/2023. Midway through procedure, Isoflurane turned on to treat hypertension. Isoflurane bracket was resting on a sidebar of the HLM. This upward lift broke the seal in the isoflurane, completely cutting gas supply to the oxygenator.

DETAILS OF INJURY TO PATIENT

Patient died post up due to complications.

TYPE OF DEVICE/MANUFACTURER

Datex Ohmeda

ACTION TAKEN

In-line monitoring picked up the problem and raised multiple alarms. 2nd perfusionist came to help. As no clear problem with gas flow due to ball in Sechrist blender floating normally and gas line appeared intact, the decision was made by perfusionist 2, to change oxygenator. Perfusionist 3 came to help with change-out. On recommencing bypass, hypoxia remained, at which point perfusionist 4 removed the vaporiser from the bracket and the issue resolved.

RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION

Full Patient Level Safety Report was carried out. On investigation,  nothing was found to be wrong with the vaporiser or the bracket.  

This is a design fault with the bracket made by G.A.S (General  Anaesthetic Services). I understand this does not have CE marking as I have tried to buy replacements, but cannot do so. MHRA not contacted as there was  no fault found.

 

 

 

Presented this case at the safety committee session at the annual  congress in Nottingham 2025.

 

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
050
1/10/19
Venous Cannula Damaged
REPORT NO
049
1/12/19
Oxygenation Issue
REPORT NO
048
1/8/19
Roller Pump Failure
REPORT NO
047
1/12/19
Broken Cannula Dilator
REPORT NO
046
6/12/19
Gas Blender Error
REPORT NO
045
20/11/19
Tubing Detachments
REPORT NO
044
27/11/19
Interruption to IABP counterpulsation
REPORT NO
043
1/7/19
Getinge Mal-Position Of IABP Catheter
REPORT NO
042
1/2/19
Avalon Dual Lumen Cannula Crack
REPORT NO
041
1/2/19
Custodial Rubber Fragmentation In Spiked Bag
REPORT NO
040
1/10/18
Reservoir Clot
REPORT NO
039
1/10/18
Pall Clot In Cardioplegia Circuit
REPORT NO
038
1/10/18
Maquet HL20 Cardioplegia Pump Error
REPORT NO
037
1/10/17
Temed D200 Haemoconcentrator Clot
REPORT NO
036
1/10/17
Oxygenator issue
REPORT NO
035
1/10/17
Medtronic Fusion Oxygenator Blood To Gas Leak
REPORT NO
034
Possible Medtronic Fusion Oxygenator Failure
REPORT NO
033
Fannin Heparin Ampules Shattering
REPORT NO
032
Sucker Placed In Raceway In Wrong Direction
REPORT NO
031
High Transmembrane Pressure Gradients
REPORT NO
030
Faulty Connector On ECMO Soft Shell Reservoir Bag
REPORT NO
029
Sorin Inspire 8F Blood Leak From Temperature Port
REPORT NO
028
Sorin Inspire 8F Oxygenator Detached From Holding Arm
REPORT NO
027
Paragon Oxygenator Gas Inlet Leak
REPORT NO
026
Maquet IABP Console Failure
REPORT NO
025
Sorin EOS Gas To Blood Leak Whilst Priming
REPORT NO
024
Chalice Paragon Midi Heat Exchanger Leak Whilst Priming
REPORT NO
023
Chalice Paragon Heat Exchanger Water To Blood Leak
REPORT NO
022
Vent Pump Boot Tubing Inserted In Wrong Way Round
REPORT NO
021
Paragon Oxygenator Heat Exchanger Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
140
139
138
137
136
135
134
133
132
131
130
129
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.