PERFUSION REPORTS

REPORT
135
DATE
3/2/2026
TITLE
Human Factors
DETAILS OF INCIDENT

During an elective AVR for severe AS, a locum perfusionist was supervising our trainee perfusionist. Bypass was initiated, root vent cannula was placed, and the cross clamp was applied. Cardioplegia delivery was initiated antegradely with cold oxygenated blood cardioplegia. The heart was not arresting, therefore, the aorta was opened and cold oxygenated blood cardioplegia was delivered down the ostia.

DETAILS OF INJURY TO PATIENT

No harm to patient.

TYPE OF DEVICE/MANUFACTURER
ACTION TAKEN

Help was called (2nd perfusionist) to investigate as the locum perfusionist was concerned that the cardioplegia was not effective. After 6 minutes and a total of 1L cold oxygenated cardioplegia the 2nd perfusionist checked the circuit and noticed that the circuit was patent and air free, however, the white clip for the cardioplegia infusion was closed off, therefore the previous delivery had been of cold oxygenated blood only. The clip was subsequently opened and 1L of cold blood cardioplegia was delivered down the ostia and the heart was successfully arrested.

The surgeon and anaesthetist were immediately informed during the fault finding and the issue was resolved. An ABG was done after cardioplegia delivery which confirmed expected normal ABG values for CPB. The rest of the case was uneventful, when the clamp was removed the heart restarted without issue and the ABGs were all as expected.

RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION

‍Staffing did not allow for the senior perfusionist on duty to supervise the trainee as they would normally, as 2 other members of staff were off. The senior perfusionist was on call and would be doing the second case. The staff perfusionist cannot supervise trainees due to departmental educational rules. Therefore, the locum perfusionist (+20 years of experience) was deemed appropriate to supervise the trainee. As a team, there could have perhaps been a second check of the pump since the locum perfusionist is new to the department.

OUTCOME ACTIONS:

  • Review of pre CPB checklist to perhaps include 'cardioplegia clip opened' or completely remove the white clip from the circuit design.
  • Perhaps new practice rule i.e. new locum perfusionists not to supervise trainees unless they have been here more than once.
  • 2 person pre CPB checklist for locum staff or new staff

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
047
1/12/19
Broken Cannula Dilator
REPORT NO
046
6/12/19
Gas Blender Error
REPORT NO
045
20/11/19
Tubing Detachments
REPORT NO
044
27/11/19
Interruption to IABP counterpulsation
REPORT NO
043
1/7/19
Getinge Mal-Position Of IABP Catheter
REPORT NO
042
1/2/19
Avalon Dual Lumen Cannula Crack
REPORT NO
041
1/2/19
Custodial Rubber Fragmentation In Spiked Bag
REPORT NO
040
1/10/18
Reservoir Clot
REPORT NO
039
1/10/18
Pall Clot In Cardioplegia Circuit
REPORT NO
038
1/10/18
Maquet HL20 Cardioplegia Pump Error
REPORT NO
037
1/10/17
Temed D200 Haemoconcentrator Clot
REPORT NO
036
1/10/17
Oxygenator issue
REPORT NO
035
1/10/17
Medtronic Fusion Oxygenator Blood To Gas Leak
REPORT NO
034
Possible Medtronic Fusion Oxygenator Failure
REPORT NO
033
Fannin Heparin Ampules Shattering
REPORT NO
032
Sucker Placed In Raceway In Wrong Direction
REPORT NO
031
High Transmembrane Pressure Gradients
REPORT NO
030
Faulty Connector On ECMO Soft Shell Reservoir Bag
REPORT NO
029
Sorin Inspire 8F Blood Leak From Temperature Port
REPORT NO
028
Sorin Inspire 8F Oxygenator Detached From Holding Arm
REPORT NO
027
Paragon Oxygenator Gas Inlet Leak
REPORT NO
026
Maquet IABP Console Failure
REPORT NO
025
Sorin EOS Gas To Blood Leak Whilst Priming
REPORT NO
024
Chalice Paragon Midi Heat Exchanger Leak Whilst Priming
REPORT NO
023
Chalice Paragon Heat Exchanger Water To Blood Leak
REPORT NO
022
Vent Pump Boot Tubing Inserted In Wrong Way Round
REPORT NO
021
Paragon Oxygenator Heat Exchanger Leak
REPORT NO
020
S3 E62 Pump Error
REPORT NO
019
Medtronic Potential Oxygenator Leak
REPORT NO
018
Medos Heat Exchanger Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
137
136
135
134
133
132
131
130
129
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.