PERFUSION REPORTS

REPORT
130
DATE
19/5/2025
TITLE
Hardware: Level Alarm
DETAILS OF INCIDENT

Patient ready to go on cardiopulmonary bypass. Heart lung machine which had passed all safety checks was already connected to the patient. Level alarm (critical alarm) had a sensor malfunction. Despite numerous attempts to reset the alarm to allow bypass to continue, the malfunction persisted.

DETAILS OF INJURY TO PATIENT

A decision was made to disconnect patient from HLM and use an alternative machine. Consultant surgeon and consultant anaesthetist informed, and new HLM and circuit obtained from perfusion, new circuit primed, checked and connected to patient and successfully placed on bypass. There was a delay of approximately 15 minutes. Patient was stable throughout.

TYPE OF DEVICE/MANUFACTURER

Spectrum contacted and Datex report submitted.

Model: Serial No: QLV1000397

ACTION TAKEN

A decision was made to disconnect patient from HLM and use an alternative machine. Consultant surgeon and consultant anaesthetist informed, and new HLM and circuit obtained from perfusion, new circuit primed, checked and connected to patient and successfully placed on bypass. There was a delay of approximately 15 minutes. Patient was stable throughout.

RESPONSE FROM MANUFACTURER

The result of the investigation is that it wascaused by suspected damaged wiring or solder connection going into or insidethe red protected level sensor which caused the fault.

ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO

https://eu-files.jotform.com/jufs/Perfusionadmin/91872635781367/6235366567124154539/d650d6bd-63aa-468e-a126-f422f7e26912.jpeg?md5=QfTpwfFwTugY8_gIOFpwzQ&expires=1761920854

SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
043
1/7/19
Getinge Mal-Position Of IABP Catheter
REPORT NO
042
1/2/19
Avalon Dual Lumen Cannula Crack
REPORT NO
041
1/2/19
Custodial Rubber Fragmentation In Spiked Bag
REPORT NO
040
1/10/18
Reservoir Clot
REPORT NO
039
1/10/18
Pall Clot In Cardioplegia Circuit
REPORT NO
038
1/10/18
Maquet HL20 Cardioplegia Pump Error
REPORT NO
037
1/10/17
Temed D200 Haemoconcentrator Clot
REPORT NO
036
1/10/17
Oxygenator issue
REPORT NO
035
1/10/17
Medtronic Fusion Oxygenator Blood To Gas Leak
REPORT NO
034
Possible Medtronic Fusion Oxygenator Failure
REPORT NO
033
Fannin Heparin Ampules Shattering
REPORT NO
032
Sucker Placed In Raceway In Wrong Direction
REPORT NO
031
High Transmembrane Pressure Gradients
REPORT NO
030
Faulty Connector On ECMO Soft Shell Reservoir Bag
REPORT NO
029
Sorin Inspire 8F Blood Leak From Temperature Port
REPORT NO
028
Sorin Inspire 8F Oxygenator Detached From Holding Arm
REPORT NO
027
Paragon Oxygenator Gas Inlet Leak
REPORT NO
026
Maquet IABP Console Failure
REPORT NO
025
Sorin EOS Gas To Blood Leak Whilst Priming
REPORT NO
024
Chalice Paragon Midi Heat Exchanger Leak Whilst Priming
REPORT NO
023
Chalice Paragon Heat Exchanger Water To Blood Leak
REPORT NO
022
Vent Pump Boot Tubing Inserted In Wrong Way Round
REPORT NO
021
Paragon Oxygenator Heat Exchanger Leak
REPORT NO
020
S3 E62 Pump Error
REPORT NO
019
Medtronic Potential Oxygenator Leak
REPORT NO
018
Medos Heat Exchanger Leak
REPORT NO
017
Avant D903 Has A Dual Chamber Reservoir Leak
REPORT NO
016
Additional Sucker Incorrectly Placed In Raceway
REPORT NO
015
Large Perished Tear Observed On Tubing At Back Of Vaporiser
REPORT NO
014
Gas Blender Flow Control Immobilised
Perfusion Report cards are shown in batches of 30 at any one time where available.
133
132
131
130
129
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.