PERFUSION REPORTS

REPORT
128
DATE
25/5/2025
TITLE
Single Station Selectatec Type Backbar for Vapouriser. 9000-001
DETAILS OF INCIDENT

It involved a SingleStation Selectatec Type Backbar, a mounting unit for Sevoforane Vapourisers. It was custom built for us by a UK Company, supplied via an Irish representative company.

No vapour was being delivered to the patient while on Bypass. The Anaesthetic registrar noticed the BIS number climbing and questioned the level of anaesthesia. He reverted to IV sedation when he became suspicious. Ironically, it was team changeover time of year for anaesthtic registratrs and an inexperienced registrar was thought to be the cause. However, when reported the Hospital engineer thought there was an internal problem. The engineer rigged the gas sampling line of the Anaesthetic monitor to the Oxygenator. That normally displays the percentage of vapour that the patient is receiving from the Anaesthetic via the ventilator, and noted no vapour. Further review noted that the vaporiser valve was fitted incorrecty and worked in reverse. So, as you increased Percentage vapour, the valve closes accordingly, not opens. This was very evident with theGas sampling line.  

These were 4 new Backbars ,built for us by the company outsourced the build to a UK company. Hence their email, attached to the Safety report.

The concern is that otherunits may have had these built from the same company and have not yet discovered the problem.

 

DETAILS OF INJURY TO PATIENT
TYPE OF DEVICE/MANUFACTURER
ACTION TAKEN

Reported to the Society and the Health Products Regulatory Authority (HPRA) Ireland. In accordance to the regulatory obligations under MDR (EU 2017/745) which requires medical devices distributors to conduct post market surveillance and report safety and performance issues. Oxygen care informed

RESPONSE FROM MANUFACTURER

Awaiting response.

ADDTIONAL INFORMATION

Connect the gas sampling line of the Anaesthetic monitor to the gas line of the HLM beforethe patient enters the Theatre to show correct Vapour delivery. This is now part of a regular check off.

 

SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
008
1/7/11
Aorta Remained Clamped Post CPB
REPORT NO
007
1/5/11
Deficiencies In The QC At Terumo Ann Arbor, USA.
REPORT NO
006
1/4/11
Awareness Of Suckers When Using Bio-Glues
REPORT NO
005
1/2/11
Patient Exsanguinated From Additional Venous Line Post CPB
REPORT NO
004
1/1/11
Possible Inaccuracies In Colloid Research
REPORT NO
003
1/2/10
Difficulties In Visualising Blender Reading
REPORT NO
002
1/1/10
Luer Cap Found In Venous Inlet Port
REPORT NO
001
1/1/09
Oxygenator Water To Blood Leak
Perfusion Report cards are shown in batches of 30 at any one time where available.
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
044
043
042
041
040
039
038
037
036
035
034
033
032
031
030
029
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.