Isoflurane vaporiser
DETAILS OF INCIDENT

Upon initiation of bypass, the perfusionist became aware of an oxygenation issue; arterial blood was dark and desaturated. Team informed, lungs switched on, came off bypass and called for help. Hardware sweep yielded no discrepancies, and air was flowing as expected - the vaporiser was set to 0% as CPB had been terminated. When turning the isoflurane to 1% it was discovered that air flow was blocked post vaporiser. Decision was made to replace the vaporiser and re-test of bypass. Air flow was now restored at 1% iso.

Bypass was initiated for the second time. There was adequate oxygen delivery, although settings were higher than expected - this was attributed to oxygen debt. Sweep and FiO2 settings came down throughout the case, consistent with this theory, although settings were still higher than usual.

Approximately an hour into the second bypass run, there was a sudden decrease in pO2 and sharp increase in pCO2 observed through CDI inline monitoring - in the space of 20 seconds. Help called again, max settings set on sweep and FiO2 and rise in pCO2, oxy recirculation line opened. This attenuated the fall in pO2 and rise in pCO2, but numbers were not recovering to pre-event levels. Decision made by 3 perfusionists to bypass the vaporiser using 1/4" connector to connect the pre and post vaporiser gas lines together. Anaesthetist started propofol infusion.

Numbers restored to safe and expected levels. CPB continued without further issues.

DETAILS OF INJURY TO PATIENT

Patient unharmed

TYPE OF DEVICE/MANUFACTURER

Ohmeda isoflurane vaporiser

ACTION TAKEN

Vaporiser was sent to medical engineering for testing. No issues were discovered. Perfusionist routine changed - previously air flow was tested with vaporiser on 0% during priming. Now, vaporiser is set to 1% when testing air flow when priming.

RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
054
1/4/20
Leaking ALF
REPORT NO
053
1/2/20
Arterial Line Filter Leak
REPORT NO
052
1/3/19
Gas Line Filter Issue
REPORT NO
051
1/12/19
Dual Lumen Cannula Movement
REPORT NO
050
1/10/19
Venous Cannula Damaged
REPORT NO
049
1/12/19
Oxygenation Issue
REPORT NO
048
1/8/19
Roller Pump Failure
REPORT NO
047
1/12/19
Broken Cannula Dilator
REPORT NO
046
6/12/19
Gas Blender Error
REPORT NO
045
20/11/19
Tubing Detachments
REPORT NO
044
27/11/19
Interruption to IABP counterpulsation
REPORT NO
043
1/7/19
Getinge Mal-Position Of IABP Catheter
REPORT NO
042
1/2/19
Avalon Dual Lumen Cannula Crack
REPORT NO
041
1/2/19
Custodial Rubber Fragmentation In Spiked Bag
REPORT NO
040
1/10/18
Reservoir Clot
REPORT NO
039
1/10/18
Pall Clot In Cardioplegia Circuit
REPORT NO
038
1/10/18
Maquet HL20 Cardioplegia Pump Error
REPORT NO
037
1/10/17
Temed D200 Haemoconcentrator Clot
REPORT NO
036
1/10/17
Oxygenator issue
REPORT NO
035
1/10/17
Medtronic Fusion Oxygenator Blood To Gas Leak
REPORT NO
034
Possible Medtronic Fusion Oxygenator Failure
REPORT NO
033
Fannin Heparin Ampules Shattering
REPORT NO
032
Sucker Placed In Raceway In Wrong Direction
REPORT NO
031
High Transmembrane Pressure Gradients
REPORT NO
030
Faulty Connector On ECMO Soft Shell Reservoir Bag
REPORT NO
029
Sorin Inspire 8F Blood Leak From Temperature Port
REPORT NO
028
Sorin Inspire 8F Oxygenator Detached From Holding Arm
REPORT NO
027
Paragon Oxygenator Gas Inlet Leak
REPORT NO
026
Maquet IABP Console Failure
REPORT NO
025
Sorin EOS Gas To Blood Leak Whilst Priming
Perfusion Report cards are shown in batches of 30 at any one time where available.
144
143
142
141
140
139
138
137
136
135
134
133
132
131
130
129
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
051
050
049
048
047
046
045
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.