PERFUSION REPORTS

REPORT
150
DATE
21/10/2026
TITLE
Governance issue of undocumented IABP therapy
DETAILS OF INCIDENT

A 75-year-old man underwent an aortic valve replacement and CABG. The operation was complex and an intra-aortic balloon pump (IABP) in situ. Over the following hours, the balloon pump intermittently alarmed for gas leak. Reviews and troubleshooting from anaesthetic, surgical, and perfusion teams occurred, and the IABP was removed during a subsequent chest washout. Shortly after removal, the patient was found to have signs of acute right leg limb ischaemia. A CT scan confirmed gas embolism in the right iliac and distal vessels. He later required a right lower limb amputation.

DETAILS OF INJURY TO PATIENT

Lower leg amputation to patient.

TYPE OF DEVICE/MANUFACTURER
ACTION TAKEN
RESPONSE FROM MANUFACTURER
ADDTIONAL INFORMATION

What happened?

‍

  • The IABP was inserted intraoperatively, the following day, the IABP console began intermittently alarming indicating a possible gas leak.

‍

  • This was escalated and was reviewed at the bedside by the perfusionist. Standard checks were conducted by the perfusionist, and no additional evidence of a significant leak/ rupture was noted. A second perfusionist reviewed later that evening.

‍

  • The perfusionist did raise with the clinical team whether the IABP was still required.

‍

  • A second perfusionist also conducted trouble shooting and swapped out the console in case that was the cause. This appeared to resolve the alarm issue at the time.

‍

  • Overnight, this IABP console continued to alarm.

‍

  • During morning review with perfusion, surgical and anaesthetic teams, blood was noted in the line at this stage for the first time, indicating possible rupture of the IABP balloon, and a decision was taken to remove the IABP.

‍

  • Removal of the IAPB was noted to be straightforward. Nil concerns noted at the time of removal. Pressure applied to the area following removal. The balloon tip was discarded and not inspected.

‍

What was expected?

‍

  • Clear documentation of all care and intervention delivered to the patient.

‍

Why was there a difference?

‍

  • Although troubleshooting appears to have been undertaken appropriately (console swap, inspection, review by on‑call perfusion), there is no clear standardised protocol for managing recurrent IABP alarms.

‍

  • Perfusionists do not document actions in the patient’s clinical notes (Cerner electronic record) or any other location, which limited visibility of troubleshooting, communication between the multi-professional team, and possibly reduced continuity of care.
SUPPORTING VISUALS/VIDEO
SUPPORTING DOCUMENT

OTHER PERFUSION
 REPORTS

London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Select and scroll to view chosen report
REPORT NO
031
High Transmembrane Pressure Gradients
REPORT NO
030
Faulty Connector On ECMO Soft Shell Reservoir Bag
REPORT NO
029
Sorin Inspire 8F Blood Leak From Temperature Port
REPORT NO
028
Sorin Inspire 8F Oxygenator Detached From Holding Arm
REPORT NO
027
Paragon Oxygenator Gas Inlet Leak
REPORT NO
026
Maquet IABP Console Failure
REPORT NO
025
Sorin EOS Gas To Blood Leak Whilst Priming
REPORT NO
024
Chalice Paragon Midi Heat Exchanger Leak Whilst Priming
REPORT NO
023
Chalice Paragon Heat Exchanger Water To Blood Leak
REPORT NO
022
Vent Pump Boot Tubing Inserted In Wrong Way Round
REPORT NO
021
Paragon Oxygenator Heat Exchanger Leak
REPORT NO
020
S3 E62 Pump Error
REPORT NO
019
Medtronic Potential Oxygenator Leak
REPORT NO
018
Medos Heat Exchanger Leak
REPORT NO
017
Avant D903 Has A Dual Chamber Reservoir Leak
REPORT NO
016
Additional Sucker Incorrectly Placed In Raceway
REPORT NO
015
Large Perished Tear Observed On Tubing At Back Of Vaporiser
REPORT NO
014
Gas Blender Flow Control Immobilised
REPORT NO
013
Isoflurane Vaporiser Seal Leak Due To Malposition
REPORT NO
012
1/2/13
Hypercoagulable Disorder May Have Led To Oxygenator Failure
REPORT NO
011
1/1/13
LV Vent Placed In Raceway In Wrong Direction
REPORT NO
010
1/3/12
Sucker Placed In Raceway In Wrong Direction
REPORT NO
009
1/2/12
Similarity In Heparin And St Thomas Cardioplegia Ampules
REPORT NO
008
1/7/11
Aorta Remained Clamped Post CPB
REPORT NO
007
1/5/11
Deficiencies In The QC At Terumo Ann Arbor, USA.
REPORT NO
006
1/4/11
Awareness Of Suckers When Using Bio-Glues
REPORT NO
005
1/2/11
Patient Exsanguinated From Additional Venous Line Post CPB
REPORT NO
004
1/1/11
Possible Inaccuracies In Colloid Research
REPORT NO
003
1/2/10
Difficulties In Visualising Blender Reading
REPORT NO
002
1/1/10
Luer Cap Found In Venous Inlet Port
Perfusion Report cards are shown in batches of 30 at any one time where available.
151
150
149
148
147
146
145
144
143
142
141
140
139
138
137
136
135
134
133
132
131
130
129
128
127
126
125
124
123
122
121
120
119
118
117
116
115
114
113
112
111
110
109
108
107
106
105
104
103
102
101
100
099
098
097
096
095
094
093
092
091
090
089
088
087
086
085
084
083
082
081
080
079
078
077
076
075
074
073
072
071
070
069
068
067
066
065
064
063
062
061
060
059
058
057
056
055
054
053
052
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
Know the report number? Scroll horizontally through the circle icons and select to view the appropriate report.